
General
Purpose
Oscilloscope
Menus
Dene
Measure
Menu
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
user
defined
Y
ou
can
set
the
unit
of
measure
for
the
upper
,
middle
,
and
lower
thresholds
to
either
%
(percent)
or
V
olts
.
P
ercent
is
calculated
from
the
top-base
values
,
and
you
can
set
the
percent
values
from
025
percent
to
+125
percent
in
0.1
percent
increments
.
V
olts
lets
you
set
the
thresholds
to
particular
voltage
values
regardless
of
the
top-base
values
.
Y
ou
can
set
the
voltage
values
from
01.00
GV
to
+1.00
GV
in
increments
determined
by
the
voltage
range
you
are
in.
F
or
example
,
if
the
upper
threshold
is
set
to
110
V
,
you
can
change
the
threshold
in
1
V
increments
.
However
,
if
the
threshold
is
set
to
110
MV
,
you
can
change
the
threshold
in
1
MV
increments
.
The
upper
threshold
value
is
always
greater
than
the
value
of
the
middle
threshold,
and
the
middle
threshold
value
is
always
greater
than
the
value
of
the
lower
threshold.
The
instrument
will
not
allow
a
threshold
to
cross
over
the
adjacent
threshold.
If
the
threshold
you
are
using
for
the
measurement
is
not
positioned
on
the
waveform,
then
a
Thresholds?
error
message
is
displayed
in
place
of
the
automatic
measurement
result.
When
Units
is
set
to
V
olts
,
user
dened
simplies
the
threshold
detection
algorithm.
The
result
is
that
the
measurement
throughput
of
the
instrument
is
increased,
because
the
instrument
does
not
have
to
calculate
the
voltage
thresholds
.
K
ey
Path
4
Dene
meas
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Thresholds...
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
user
defined
10-10
Summary of Contents for 54750A
Page 1: ...User s Guide HP 83480A Analyzer HP 54750A Oscilloscope ...
Page 6: ...NOTE Clean the cabinet using a damp cloth only vi ...
Page 7: ...X Ray Radiation Notice vii ...
Page 8: ...Declaration of Conformity viii ...
Page 17: ...Figure0 1 Exampleof astatic safeworkstation xvii ...
Page 28: ......
Page 43: ...1 The Instrument at a Glance ...
Page 57: ...The Rear Panel Figure1 3 Theinstrument rearpanel 1 15 ...
Page 60: ...TheInstrument ataGlance ...
Page 61: ...2 General Purpose Keys ...
Page 69: ...3 Speci cations and Characteristics ...
Page 76: ...Speci cationsand Characteristics ...
Page 77: ...4 Calibration Overview ...
Page 84: ...CalibrationOverview ...
Page 85: ...5 Eye Mask and Eyeline Mode Measurement Tutorials ...
Page 136: ...Eye Mask andEyelineModeMeasurement Tutorials ...
Page 137: ...6 The Digital Communications Analysis Menus ...
Page 173: ...The Digital Communications Analysis Menus MeasureEyeMenu Figure6 7 Crossing measurement 6 37 ...
Page 178: ...The Digital Communications Analysis Menus MeasureEye Menu Figure6 11 MeasuringQ factor 6 42 ...
Page 191: ...7 Waveform Measurements ...
Page 211: ...8 Making Automatic Measurements ...
Page 233: ...9 Increasing Measurement Accuracy and Time Interval Measurement ...
Page 252: ...IncreasingMeasurement Accuracy andTime Interval Measurement ...
Page 253: ...10 General Purpose Oscilloscope Menus ...
Page 317: ...11 The General Function Menus ...
Page 345: ...The General Function Menus DisplayMenu Figure11 3 Connecteddots 11 29 ...
Page 415: ...12 Messages ...
Page 421: ...13 How the Instrument Works ...
Page 453: ...Index ...