
General
Purpose
Oscilloscope
Menus
FFT
Menu
the
sample
frequency
.
The
resolution
or
the
spacing
between
the
points
in
the
frequency
record
is
Fs/N.
The
frequency
Fs/2
is
a
unique
frequency
referred
to
as
the
Nyquist
frequency
.
At
the
Nyquist
frequency
there
are
exactly
two
samples
on
every
cycle
of
the
input
signal.
Signals
above
the
Nyquist
frequency
become
aliased
which
means
that
they
appear
as
signals
of
a
lower
frequency
.
The
reason
is
that
there
are
not
enough
sample
points
on
each
cycle
of
the
signal
to
determine
the
correct
frequency
.
It
turns
out
that
the
points
above
Fs/2
are
a
mirror
image
of
the
points
below
Fs/2.
The
points
above
Fs/2
are
not
displayed
because
they
do
not
provide
any
additional
information.
Therefore
,
N
time
samples
results
in
N/2
displayed
frequency
points
.
Figure
10-3.
Sine
wave
in
the
frequency
domain.
Figure
10-4
shows
an
example
.
The
upper
graticule
shows
a
1
GHz
waveform
in
the
time
domain,
and
the
lower
graticule
shows
the
same
signal
in
the
frequency
domain.
By
looking
at
the
Magnify
Span
and
Center
Freq
softkeys
,
you
can
determine
the
display
setup
and
see
there
is
strong
signal
content
at
1
GHz ,
as
well
as
odd
frequency
multiples
(3
GHz,
5
GHz,
and
so
on).
10-29
Summary of Contents for 54750A
Page 1: ...User s Guide HP 83480A Analyzer HP 54750A Oscilloscope ...
Page 6: ...NOTE Clean the cabinet using a damp cloth only vi ...
Page 7: ...X Ray Radiation Notice vii ...
Page 8: ...Declaration of Conformity viii ...
Page 17: ...Figure0 1 Exampleof astatic safeworkstation xvii ...
Page 28: ......
Page 43: ...1 The Instrument at a Glance ...
Page 57: ...The Rear Panel Figure1 3 Theinstrument rearpanel 1 15 ...
Page 60: ...TheInstrument ataGlance ...
Page 61: ...2 General Purpose Keys ...
Page 69: ...3 Speci cations and Characteristics ...
Page 76: ...Speci cationsand Characteristics ...
Page 77: ...4 Calibration Overview ...
Page 84: ...CalibrationOverview ...
Page 85: ...5 Eye Mask and Eyeline Mode Measurement Tutorials ...
Page 136: ...Eye Mask andEyelineModeMeasurement Tutorials ...
Page 137: ...6 The Digital Communications Analysis Menus ...
Page 173: ...The Digital Communications Analysis Menus MeasureEyeMenu Figure6 7 Crossing measurement 6 37 ...
Page 178: ...The Digital Communications Analysis Menus MeasureEye Menu Figure6 11 MeasuringQ factor 6 42 ...
Page 191: ...7 Waveform Measurements ...
Page 211: ...8 Making Automatic Measurements ...
Page 233: ...9 Increasing Measurement Accuracy and Time Interval Measurement ...
Page 252: ...IncreasingMeasurement Accuracy andTime Interval Measurement ...
Page 253: ...10 General Purpose Oscilloscope Menus ...
Page 317: ...11 The General Function Menus ...
Page 345: ...The General Function Menus DisplayMenu Figure11 3 Connecteddots 11 29 ...
Page 415: ...12 Messages ...
Page 421: ...13 How the Instrument Works ...
Page 453: ...Index ...