ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
6.7
6
U
SER
I
NTERFACE
When a slide cassette is placed into the instrument, the system assumes it is “full” until it checks for
the presence of slides in the cassette, which it does when first attempting to retrieve slides from the
cassette. The status is “partial” if the cassette is not filled to the top or after the system has used slides
out of a cassette. The status is “empty” when the cassette has no slides left. If a slide cassette is not
present, the status is “no cassette.”
Touch the button for each cassette bay to display the selection of slide types. Select the exact type by
touching the name on the screen. The button label for that cassette will change to reflect the selection.
The types of microscope slides are:
•
Gyn
- ThinPrep microscope slides for gynecologic cytology
•
Imager
- slides for use on the ThinPrep Imaging System (gynecologic sample); they bear
fiducial marks required by the Imager
•
Non-Gyn
- microscope slides for general cytology
•
UroCyte
- microscope slides for use in conjunction with the Vysis
®
UroVysion assay
When a cassette is empty, the system will draw off of the next cassette that has the same type of slides
in it. If no other cassette has the same type of slides, then the user is notified that there are no slides
available.
Important:
The system can sense whether slides are in a cassette or not, but it cannot sense
what type of slide it is. The system only knows what slide type was set previously. The user must
always select the slide type.
Summary of Contents for ThinPrep 5000
Page 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Page 5: ...This page intentionally left blank...
Page 6: ...Instructions For Use Instructions For Use...
Page 43: ...Table of Contents Table of Contents...
Page 48: ...1 Introduction 1 Introduction...
Page 73: ...2 Installation 2 Installation...
Page 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Page 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Page 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Page 134: ...6 User Interface 6 User Interface...
Page 217: ...7 Operating Instructions 7 Operating Instructions...
Page 256: ...8 Maintenance 8 Maintenance...
Page 283: ...9 Troubleshooting 9 Troubleshooting...
Page 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Page 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Page 316: ...Service Information Service Information...
Page 319: ...Ordering Information Ordering Information...
Page 326: ...Index Index...
Page 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...