7.22
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
7
Slide Labeling Requirements
Slides that are processed directly from the carousel must be pre-labeled, as the system does not laser
etch these samples. Note that some slides may already be laser etched but did not process due to a
sample error. These may be used as they are.
Other slides must bear a label with an accession ID that matches the ID on the vial. (Refer to
Advanced Processing Options, page 7.28 for disabling slide ID match temporarily.)
Slide Barcode Label Format
Barcode labels must be 1-D or 2-D and may have alphanumeric character strings of 5 to 64 characters
with human readable accession ID. Slide labels may be printed and applied or directly printed or
etched onto the slide, but make sure the contrast is sufficient for the scanner to read the label.
The barcode label must be one of the symbologies supported:
•
1-D Code 128
•
1-D EAN-13/JAN
•
1-D Codabar (NW7)
•
1-D Interleaved 2 of 5
•
1-D Code 39
Table 7.3: Status Indicators
CAROUSEL
DOORS
BATHS
WASTE
POWER
Status OK, ready to
process
Status OK, ready to
process
Status OK, ready to
process
Status OK, ready to process
Status OK, ready to
process
Carousel not
detected. Insert car-
ousel or make sure it
is in position
One or more doors
are open. Close the
doors.
A fixative bath is not
detected. Insert a
fixative bath and
close the door.
Liquid waste must be
emptied. See page 8.10.
The system is running
on battery power
(UPS). If a batch is in
process, it will finish
the sample and pause
the batch.
The status of the
carousel is unknown
when the door is
open.
The UPS is not
detected
or
battery is low in
power.
SMALL BATCH MODE
Summary of Contents for ThinPrep 5000
Page 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Page 5: ...This page intentionally left blank...
Page 6: ...Instructions For Use Instructions For Use...
Page 43: ...Table of Contents Table of Contents...
Page 48: ...1 Introduction 1 Introduction...
Page 73: ...2 Installation 2 Installation...
Page 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Page 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Page 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Page 134: ...6 User Interface 6 User Interface...
Page 217: ...7 Operating Instructions 7 Operating Instructions...
Page 256: ...8 Maintenance 8 Maintenance...
Page 283: ...9 Troubleshooting 9 Troubleshooting...
Page 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Page 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Page 316: ...Service Information Service Information...
Page 319: ...Ordering Information Ordering Information...
Page 326: ...Index Index...
Page 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...