ThinPrep™ 5000 System
Instructions for Use
English
AW-22289-001 Rev. 003 11-2021
30/36
Between-Instrument Precision
The study was designed to examine the ability of the ThinPrep 5000 system to prepare
reproducible slides from the same patient specimen using multiple instruments. A total of 120
specimens were enrolled in the study. Each specimen was split into three portions and
processed on three instruments. The slides were stained, coverslipped, and then reviewed by
cytotechnologists. The resulting diagnoses and specimen adequacy determinates are presented
below. One hundred seventeen (117) specimens had all three satisfactory ThinPrep 5000 slides,
one specimen had two slides with UNSAT result and one slide with Satisfactory result, one
specimen had two slides with Satisfactory result and one slide with UNSAT result, and one
specimen was excluded from analysis due to a broken slide. For comparison, the same
procedure was carried out using a ThinPrep 2000 system, with results also presented below.
Table 29: Between-Instrument Precision
ThinPrep 5000
ThinPrep 2000*
Percent of specimens that have three matching NILM
replicates or three matching ASC-US+ replicates
94.0%
(110/117)
(88.2% to 97.1%)
91.1%
(102/112)
(84.3% to 95.1%)
Percent of specimens that have three matching <LSIL
replicates or three matching LSIL+ replicates
97.4%
(114/117)
(92.7% to 99.1%)
94.6%
(106/112)
(88.8% to 97.5%)
Percent of specimens that have three matching <HSIL
replicates or three matching HSIL+ replicates
98.3%
(115/117)
(94.0% to 99.5%)
100%
(112/112)
(96.7% to 100%)
Percent of specimens that have three matching
Satisfactory replicates or three matching UNSAT
replicates
98.3%
(117/119)
(94.1% to 99.5%)
98.3%
(113/115)
(93.9% to 99.5%)
* 120 specimens were enrolled, but 5 were excluded due to slide breakage and other errors
.
Cell Count Study
The quantity of cellular material transferred onto slides, comparing ThinPrep 5000 to the
ThinPrep 2000, was evaluated in a laboratory study using a split-sample technique.
Two hundred ten (210) specimens were enrolled in the study (139 NILM, 28 ASC-US, 28 LSIL, and
15 HSIL). Each specimen was split into two parts, processed on a ThinPrep 2000 and ThinPrep
Summary of Contents for ThinPrep 5000
Page 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Page 5: ...This page intentionally left blank...
Page 6: ...Instructions For Use Instructions For Use...
Page 43: ...Table of Contents Table of Contents...
Page 48: ...1 Introduction 1 Introduction...
Page 73: ...2 Installation 2 Installation...
Page 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Page 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Page 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Page 134: ...6 User Interface 6 User Interface...
Page 217: ...7 Operating Instructions 7 Operating Instructions...
Page 256: ...8 Maintenance 8 Maintenance...
Page 283: ...9 Troubleshooting 9 Troubleshooting...
Page 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Page 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Page 316: ...Service Information Service Information...
Page 319: ...Ordering Information Ordering Information...
Page 326: ...Index Index...
Page 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...