N
ON
-G
YNECOLOGIC
S
AMPLE
P
REPARATION
5.18
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
5
URINE SPECIMENS FOR THE VYSIS
®
UROVYSION ASSAY
For use with Vysis UroVysion. If performing urine cytology, follow the Body Fluids protocol.
7. Allow to stand in PreservCyt Solution for 15 minutes (page 5.11)
8. Run on ThinPrep
™
5000 processor with AutoLoader using
Sequence
Non-Gyn.
Fix, stain, and evaluate.
1. Collection. Collect urine,
OR
process urine fresh.
Note:
Fresh urine can be mixed with a 2:1 urine-to-PreservCyt
™
Solution ratio and stored for up to 48 hours before processing.
Note:
Do not exceed a 2:1 ratio of urine to PreservCyt
™
Solution. If the
urine volume exceeds 60 mL, pour off excess. A minimum
volume of 33 mL of urine is required to perform the Vysis
®
UroVysion assay.
2. Concentrate by centrifugation (page 5.5)
Transfer the sample evenly into two labeled 50-mL centrifuge tubes.
Centrifuge at 600g for 10 minutes or 1200 g for 5 minutes.
3. Pour off supernatant and resuspend cell pellet (page 5.7)
Resuspension can be done on a vortexor or may be achieved by syringing
the pellet back and forth with a plastic pipette.
Summary of Contents for ThinPrep 5000
Page 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Page 5: ...This page intentionally left blank...
Page 6: ...Instructions For Use Instructions For Use...
Page 43: ...Table of Contents Table of Contents...
Page 48: ...1 Introduction 1 Introduction...
Page 73: ...2 Installation 2 Installation...
Page 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Page 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Page 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Page 134: ...6 User Interface 6 User Interface...
Page 217: ...7 Operating Instructions 7 Operating Instructions...
Page 256: ...8 Maintenance 8 Maintenance...
Page 283: ...9 Troubleshooting 9 Troubleshooting...
Page 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Page 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Page 316: ...Service Information Service Information...
Page 319: ...Ordering Information Ordering Information...
Page 326: ...Index Index...
Page 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...