O
PERATING
I
NSTRUCTIONS
7.28
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
7
Disable Slide ID Match
Figure 7-26 Disable Slide ID Match
To run the specimen:
1. Load one vial and appropriate filter and slide type into any position on the carousel.
2. Load the carousel into the processor.
3. Put a filled fixative bath - with empty slide rack - into the bath compartment.
4. Close all doors.
5. Press the
Advanced
button on the main screen.
6. Press the
Disable Slide ID Match
setting button.
7. Select the sample type that is to be processed and press the green
OK
button.
The display returns to the main screen. Advanced is selected and the selected options are
detailed next to it. Figure 7-27.
8. Press the
Start
button to process the sample.
Disable slide ID
match
- in order
to run a sample
with the vial/slide
ID match turned
off.
Select sample
type in order to
run the correct
process
sequence
OK
button to
accept selec-
tion and return
to main screen
SMALL BATCH MODE
Summary of Contents for ThinPrep 5000
Page 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Page 5: ...This page intentionally left blank...
Page 6: ...Instructions For Use Instructions For Use...
Page 43: ...Table of Contents Table of Contents...
Page 48: ...1 Introduction 1 Introduction...
Page 73: ...2 Installation 2 Installation...
Page 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Page 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Page 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Page 134: ...6 User Interface 6 User Interface...
Page 217: ...7 Operating Instructions 7 Operating Instructions...
Page 256: ...8 Maintenance 8 Maintenance...
Page 283: ...9 Troubleshooting 9 Troubleshooting...
Page 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Page 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Page 316: ...Service Information Service Information...
Page 319: ...Ordering Information Ordering Information...
Page 326: ...Index Index...
Page 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...