N
ON
-G
YNECOLOGIC
S
AMPLE
P
REPARATION
5.22
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
5
Bloody or Proteinaceous Specimens
Problem
Procedure
A. Did the “Sample Is
Dilute” message appear
during processing?
1. Check to see if cellularity is adequate. If
not, use more of the pellet if available.
Prepare a slide using sequence Non-Gyn.
B. Does the slide have an
obvious “halo” of cellular
material and/or white blood
cells?
1. Dilute the sample 20:1. Use a calibrated
pipette to add 1 mL of sample to a new
PreservCyt
™
Solution vial. Prepare slide
using sequence Non-Gyn.
If a halo is present on the new slide, call
Hologic Technical Service (page 12.1)
C. Is the slide sparse and
does it contain blood,
protein, or non-cellular
debris?
1. Pour the contents of the PreservCyt sample
vial into a centrifuge tube.
Call Hologic Technical
Service.
2. Concentrate by centrifugation — 600g x
10 min. (page 5.5) or 1200 g for 5 min.
YES
NO
YES
NO
YES
NO
Summary of Contents for ThinPrep 5000
Page 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Page 5: ...This page intentionally left blank...
Page 6: ...Instructions For Use Instructions For Use...
Page 43: ...Table of Contents Table of Contents...
Page 48: ...1 Introduction 1 Introduction...
Page 73: ...2 Installation 2 Installation...
Page 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Page 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Page 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Page 134: ...6 User Interface 6 User Interface...
Page 217: ...7 Operating Instructions 7 Operating Instructions...
Page 256: ...8 Maintenance 8 Maintenance...
Page 283: ...9 Troubleshooting 9 Troubleshooting...
Page 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Page 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Page 316: ...Service Information Service Information...
Page 319: ...Ordering Information Ordering Information...
Page 326: ...Index Index...
Page 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...