ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
6.9
6
U
SER
I
NTERFACE
Process Sequences
Press the button for the vial tray number (1, 2, 3 or 4) and then select the process sequence for the
sample vials that will be run in that tray. Only one type of sample may be run in each vial tray.
Figure 6-8 Select Vial Tray Sequence
For running a tray of gynecologic specimens that will be
imaged with the ThinPrep Imaging System.
Use clear ThinPrep Pap test filters and ThinPrep microscope
slides for use with the ThinPrep Imaging System.
For running a tray of non-gynecologic specimens.
Use blue ThinPrep non-gynecologic filters and ThinPrep
microscope slides.
For running a tray of urine specimens for use in conjunction
with the Vysis
®
UroVysion assay.
Use yellow ThinPrep UroCyte filters and ThinPrep UroCyte
microscope slides.
For running a tray of gynecologic specimens.
Use clear ThinPrep
™
Pap test filters and ThinPrep
microscope slides.
Important:
The system can sense whether vial trays are present or not, but it cannot sense
what type of samples they hold. The system only knows what process sequence type was set
previously for that sample tray position. The user must always select the process sequence.
Summary of Contents for ThinPrep 5000
Page 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Page 5: ...This page intentionally left blank...
Page 6: ...Instructions For Use Instructions For Use...
Page 43: ...Table of Contents Table of Contents...
Page 48: ...1 Introduction 1 Introduction...
Page 73: ...2 Installation 2 Installation...
Page 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Page 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Page 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Page 134: ...6 User Interface 6 User Interface...
Page 217: ...7 Operating Instructions 7 Operating Instructions...
Page 256: ...8 Maintenance 8 Maintenance...
Page 283: ...9 Troubleshooting 9 Troubleshooting...
Page 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Page 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Page 316: ...Service Information Service Information...
Page 319: ...Ordering Information Ordering Information...
Page 326: ...Index Index...
Page 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...