3.2 Setting Measurement Conditions on the Wiring Diagram Screen
43
3
Chapter 3 Connect
ing to
Lines to be
M
easure
d
3
3
Select the current channel.
To measure multiple circuits with the
wiring, select the corresponding chan-
nel and then set the clamp sensor and
current range.
4
Select the clamp sensor.
• When measuring power lines using multiple channels, combine multiple
clamp sensor types.
For example, when measuring 3-phase/4-wire lines, use the same clamp
sensor for channels 1 to 3.
• When using the CT9667 Flexible Clamp on Sensor, use the same value for
the sensor range setting and the instrument’s clamp sensor range setting.
• When using the 9667 Flexible Clamp on Sensor, select the CT9667.
• The 9657-10 and 9675 leakage current measurement sensors can only be
selected when the wiring is set to
[I only]
.
1P2Wx2
I1, I2
1P2Wx3
I1, I2, I3
1P3W+I
I12, I3
I
I12, I3
I
I12, I3
I only x 2 (Ix2)
I1, I2
I only x 3 (Ix3)
I1, I2, I3
9660
Load current (power)
measurement sensors
9661
CT9667
CT9667-5k
9669
9694
9695-02
9695-03
9657-10
Leakage current mea-
surement sensors
9675
Summary of Contents for PW3360-20
Page 2: ......
Page 66: ...3 8 Verifying Correct Wiring Wiring Check 58 ...
Page 104: ...7 2 Settings That Can Be Added to Quick Set Settings 96 ...
Page 120: ...8 8 Formatting the SD Memory Card or Internal Memory 112 ...
Page 166: ...11 4 Outputting a Pulse Signal 158 ...
Page 210: ...13 4 Disposing of the Instrument 202 ...
Page 218: ...A8 ...
Page 222: ...Index Index 4 ...
Page 223: ......
Page 224: ......