13.1 Trouble Shooting
194
If the cause of the issue remains unclear, reset the system. Doing so will initialize
settings to their factory defaults.
See:
4.5, "Initializing the Instrument (System Reset)" (p. 74)
Measured values
do not stabilize.
• Is the frequency of the
lines being measured 50
or 60 Hz?
The instrument does not
support 400 Hz frequency
lines.
The instrument can only be used with
50/60 Hz lines. Lines operating at 400
Hz cannot be measured.
• If the wiring setting is
1P2W/1P3W/3P3W/
3P4W, is voltage input
being supplied?
The instrument may not be
able to perform stable
measurement without volt-
age input.
If not measuring voltage, select the
current-only wiring and set the fre-
quency setting to the measurement
line frequency (50/60 Hz).
See:
4.3, "Changing Recording (Save)
Settings" (p. 65)
Unable to charge
the 9459 Battery
Pack (the Charge
LED does not
light up).
• Verify that the ambient
temperature is within the
range of 10°C to 40°C.
The instrument’s battery can be
charged within the ambient tempera-
ture range of 10°C to 40°C.
See:
"Installing (replacing) the Battery
Pack" (p. 24)
• Has the instrument been
stored for an extended
period of time with the bat-
tery pack installed?
The battery pack may have
degraded, signaling that it needs to
be replaced. Please purchase a new
battery pack. Contact your Hioki dis-
tributor for more information. If the
instrument will not be used for one
month or longer, remove the battery
pack and store at -20°C to 30°C.
See:
"Installing (replacing) the Battery
Pack" (p. 24)
The battery pack
can only be used
for a short period
of time.
• The batter pack’s capacity
may have deteriorated due
to degradation.
Symptom
Check Item, or Cause
Remedy and Reference
Summary of Contents for PW3360-20
Page 2: ......
Page 66: ...3 8 Verifying Correct Wiring Wiring Check 58 ...
Page 104: ...7 2 Settings That Can Be Added to Quick Set Settings 96 ...
Page 120: ...8 8 Formatting the SD Memory Card or Internal Memory 112 ...
Page 166: ...11 4 Outputting a Pulse Signal 158 ...
Page 210: ...13 4 Disposing of the Instrument 202 ...
Page 218: ...A8 ...
Page 222: ...Index Index 4 ...
Page 223: ......
Page 224: ......