3.2 Setting Measurement Conditions on the Wiring Diagram Screen
44
5
Select the current range.
If you do not know the appropriate
range, configure the current range set-
ting while checking the current value on
the
[WIR, CHK]
screen after connect-
ing the instrument.
See:
3.7, "Setting the Current Range" (p.
52)
9660
5A, 10A, 50A, 100A
9661
5A, 10A, 50A, 100A, 500A
CT9667
500A
CT9667-5k
5kA
9669
100A, 200A, 1kA
9694
0.5A, 1A, 5A, 10A, 50A
9695-02
9695-03
5A, 10A, 50A, 100A
9657-10
50mA, 100mA, 500mA, 1A,
5A
9675