12.7 Model PW9003 Voltage Line Power Adapter
190
12.7 Model PW9003 Voltage Line Power
Adapter
Input terminal
Banana input terminal
Connect the L9438-53 voltage Cords (2) that come with the
PW3360.
Output cord
Connect the two banana leads (connect to the PW3360’s volt-
age input terminals and supply the measurement voltage) and
one AC adapter connection cord (connect to the Z1006 AC
Adapter that comes with the PW3360 and supply power).
Operating
environment
Indoors, Pollution degree 2, altitude up to 2,000 m (6562-ft.)
Rated voltage
AC240 V
Rated current
AC3.15 A
Maximum rated
voltage to earth
300 V
Measurement Categories III
(anticipated transient overvoltage 4000 V)
Dielectric strength
(50 Hz /60 Hz,
60 minutes)
4.29 kVrms (1 mA sense current)
Between power supply pins and case
Operating tempera-
ture and humidity
-10°C to 50°C (14°F to 122°F), 80%RH or less
(non-condensating)
Storage tempera-
ture and humidity
-20°C to 60°C (-4°F to 140°F), 80%RH or less
(non-condensating)
Dimensions
Case: Approx. 125W×50H×36D mm (4.92”W×1.97”H×1.42”D,
excluding protruding parts)
Cord length:
PW3360 voltage input terminal connection side; Approx. 380
mm (14.96”)
AC adapter connection side; Approx. 380 mm (14.96”)
Mass
Approx. 180 g (6.3 oz.)
Applicable
standards
Safety EN61010 Pollution degree 2
Summary of Contents for PW3360-20
Page 2: ......
Page 66: ...3 8 Verifying Correct Wiring Wiring Check 58 ...
Page 104: ...7 2 Settings That Can Be Added to Quick Set Settings 96 ...
Page 120: ...8 8 Formatting the SD Memory Card or Internal Memory 112 ...
Page 166: ...11 4 Outputting a Pulse Signal 158 ...
Page 210: ...13 4 Disposing of the Instrument 202 ...
Page 218: ...A8 ...
Page 222: ...Index Index 4 ...
Page 223: ......
Page 224: ......