12.6 Range Configuration and Accuracy by Clamp Sensor
188
*The 500.00 A range is only available with the 9661 Clamp on Sensor.
When the 9660, 9661, or 9695-03 is Used
Power ranges
Voltage
Wiring
Current range
5.0000 A
10.000 A
50.000 A
100.00 A
500.00 A
600.00 V
1P2W
3.0000 kW
6.0000 kW
30.000 kW
60.000 kW
300.00 kW
1P3W
1P3W1U
3P3W2M
3P3W3M
6.0000 kW
12.000 kW
60.000 kW
120.00 kW
600.00 kW
3P4W
9.0000 kW
18.000 kW
90.000 kW
180.00 kW
900.00 kW
Combined accuracy
Current range
Model 9660 sensor
Model 9695-03 sensor
Model 9661 sensor
500.00A
-
±0.6%rdg.±0.11%f.s.
100.00A
±0.6%rdg.±0.12%f.s.
±0.6%rdg.±0.15%f.s.
50.000A
±0.6%rdg.±0.14%f.s.
±0.6%rdg.±0.2%f.s.
10.000A
±0.6%rdg.±0.3%f.s.
±0.6%rdg.±0.6%f.s.
5.0000A
±0.6%rdg.±0.5%f.s.
±0.6%rdg.±1.1%f.s.
When the 9669 is Used
Power ranges
Voltage
Wiring
Current range
100.00 A
200.00 A
1.0000 kA
600.00 V
1P2W
60.000 kW
120.00 kW
600.00 kW
1P3W
1P3W1U
3P3W2M
3P3W3M
120.00 kW
240.00 kW
1.2000 MW
3P4W
180.00 kW
360.00 kW
1.8000 MW
Combined accuracy
Current range
Model 9669 sensor
1.0000kA
±1.3%rdg.±0.11%f.s.
200.00 A
±1.3%rdg.±0.15%f.s.
100.00 A
±1.3%rdg.±0.2%f.s.
Summary of Contents for PW3360-20
Page 2: ......
Page 66: ...3 8 Verifying Correct Wiring Wiring Check 58 ...
Page 104: ...7 2 Settings That Can Be Added to Quick Set Settings 96 ...
Page 120: ...8 8 Formatting the SD Memory Card or Internal Memory 112 ...
Page 166: ...11 4 Outputting a Pulse Signal 158 ...
Page 210: ...13 4 Disposing of the Instrument 202 ...
Page 218: ...A8 ...
Page 222: ...Index Index 4 ...
Page 223: ......
Page 224: ......