![Hioki PW3360-20 Instruction Manual Download Page 61](http://html1.mh-extra.com/html/hioki/pw3360-20/pw3360-20_instruction-manual_2137144061.webp)
3.7 Setting the Current Range
53
3
Chapter 3 Connect
ing to
Lines to be
M
easure
d
3
3
Check the current value and set
the current range.
When measuring multiple single-phase/
2-wire (1P2W) circuits or when “current
only” is selected, you can select other
channels in the same way and set the
range.
Selecting an appropriate range
Select an appropriate range based in-
formation such as the load rating, oper-
ating conditions, and breaker rating. If
the range is too low, you will experience
an over-range event during measure-
ment, making accurate measurement impossible. Conversely, if the range is too
high, the error will be too large, making accurate measurement impossible. Set the
current range based on the maximum load current that is anticipated during the
measurement interval.
9660
5A, 10A, 50A, 100A
9661
5A, 10A, 50A, 100A, 500A
CT9667
500A
CT9667-5k
5kA
9669
100A, 200A, 1kA
9694
0.5A, 1A, 5A, 10A, 50A
9695-02
9695-03
5A, 10A, 50A, 100A
9657-10
50mA, 100mA, 500mA, 1A, 5A
9675
The current range setting can be configured on the Measurement
screen,
[SET 1/8, MEAS 1]
screen, or Wirings screen.
[MEAS, LIST]
screen
[SET 1/8, MEAS 1]
screen
Summary of Contents for PW3360-20
Page 2: ......
Page 66: ...3 8 Verifying Correct Wiring Wiring Check 58 ...
Page 104: ...7 2 Settings That Can Be Added to Quick Set Settings 96 ...
Page 120: ...8 8 Formatting the SD Memory Card or Internal Memory 112 ...
Page 166: ...11 4 Outputting a Pulse Signal 158 ...
Page 210: ...13 4 Disposing of the Instrument 202 ...
Page 218: ...A8 ...
Page 222: ...Index Index 4 ...
Page 223: ......
Page 224: ......