12.2 Basic Specifications
163
Chapter 12 S
p
ecif
icat
ions
12
A/D converter
resolution
16 bit
Display range
Voltage
: 5 V to 1,000 V; separate warning displayed when over-
range
Zero-suppression processing forces voltage RMS values
of less than 5 V to be displayed as 0 V.
Current
: 0.4% to 130% of range
Separate warning displayed when over-range or over-
peak.
Zero-suppression processing forces current RMS values
of less than 0.4% to be displayed as 0 A.
Power
: 0% to 130% of range (zero-suppression is triggered when
the voltage RMS value or current RMS value is 0)
Effective
measuring
range
Voltage
: 90 V to 780 V; peak: ±1,400 V
Current
: 5% to 110% of range; peak: ±400% of range
However, maximum range is 200%.
Power
: 5% to 110% of range
Frequency : 45 Hz to 66 Hz
Measurement
items
Voltage RMS, current RMS, voltage fundamental wave value, current
fundamental wave value, voltage fundamental wave phase angle,
current fundamental wave phase angle, frequency (U1), voltage
waveform peak (absolute value), current waveform peak (absolute
value), active power, reactive power (with lag/lead display), apparent
power, power factor (with lag/lead display) or displacement power
factor (with lag/lead display), active energy (consumption, regenera-
tion), reactive energy (lag, lead), active power demand quantity (con-
sumption, regeneration), reactive power demand quantity (lag, lead),
active power demand value (consumption, regeneration), reactive
power demand value (lag, lead), power factor demand, pulse input
Measurement specifications
Display range, effective measurement range, effective peak range chart
(representative example: 9661 sensor)
Item
Range
Display
range
Effective measurement
range
Display
range
Effective peak
Lower limit
Lower limit
Upper limit
Upper limit
Range
Voltage
600 V single range
5.00 V
90.00 V
780.00 V
1000.0 V
±
1400 Vpeak
Current
(Model 9661)
5 A range
0.0200 A
0.2500 A
5.5000 A
6.5000 A
±
20 Apeak
10 A range
0.040 A
0.500 A
11.000 A
13.000 A
±
40 Apeak
50 A range
0.200 A
2.500 A
55.000 A
65.000 A
±
200 Apeak
100 A range
0.40 A
5.00 A
110.00 A
130.00 A
±
400 Apeak
500 A range
2.00 A
25.00 A
550.00 A
650.00 A
±
1000 Apeak
Summary of Contents for PW3360-20
Page 2: ......
Page 66: ...3 8 Verifying Correct Wiring Wiring Check 58 ...
Page 104: ...7 2 Settings That Can Be Added to Quick Set Settings 96 ...
Page 120: ...8 8 Formatting the SD Memory Card or Internal Memory 112 ...
Page 166: ...11 4 Outputting a Pulse Signal 158 ...
Page 210: ...13 4 Disposing of the Instrument 202 ...
Page 218: ...A8 ...
Page 222: ...Index Index 4 ...
Page 223: ......
Page 224: ......