background image

Rev. 6 December 2017 

Copyright © CO2 Meter, Inc. All Rights Reserved. 

User’s Manual 

25%

 

O2

 

Sampling

 

Data

 

Logger

 

Model

 

CM

0505

 

Summary of Contents for CM-0505

Page 1: ...Rev 6 December 2017 Copyright CO2 Meter Inc All Rights Reserved User s Manual 25 O2 Sampling Data Logger Model CM 0505 ...

Page 2: ...ial Number Purchase Date WARNING Install our complimentary GasLab software before connecting your CO2Meter product s to your computer Failure to do so may affect the ability for GasLab to detect your sensor If this happens please follow the instructions shown in the Troubleshooting Guide on page 21 ...

Page 3: ... OPERATION GUIDE 11 CONNECTING METER 11 SETUP 12 SETTINGS 13 DISPLAY SETTINGS 13 PORT SETTINGS 13 TIME DATE SETTINGS 13 LOGGING DATA 14 FOR EACH LOGGING SESSION 14 LOGGING SETTINGS 14 QUICK START GUIDE 16 CALIBRATION 17 WHY IS CALIBRATION REQUIRED 17 FACTORY CALIBRATION 17 ORDERING A CERTIFICATE OF CALIBRATION 18 CALIBRATION PROCEDURE 18 SPECIFICATIONS 20 TROUBLESHOOTING GUIDE 21 USB DRIVER INSTAL...

Page 4: ...rst focus allows CO2Meter Inc to continue to provide solutions for the future Based in Ormond Beach FL CO2Meter Inc is committed to the success of our customers the health welfare and prosperity of our talented employees and the continued development of our local community CO2Meter Inc appreciates your business and looks forward to working with you and your team in the future Please take some time...

Page 5: ... prevent the potential for battery leakage 7 Make sure that the tubes are securely attached to the meter before sampling a closed environment 8 Do not operate with an obstructed sample path 9 This meter is not designed for outdoor use 10 Do not expose this meter to water or any liquids 11 Do not operate this meter if the enclosure is opened 12 Do not operate the device if it is malfunctioning SAVE...

Page 6: ...ng bulkhead fitting o 2 1 8 Barb 10 32 UNF fittings o 2 Particulate filter CM 0117 o 2 Hydrophobic filter CM 0118 o 1 Moisture trap CM 0112 Optional Accessories If the meter is to operate in environments where humidity is very high 95 RH an Extreme Moisture Filter CM 0103 can be purchased separately This filter will allow free flow of sampled air while keeping moisture out of the meter A 10 000 ho...

Page 7: ...run using a Windows 7 8 virtual machine software such as VMware Fusion or similar Microsoft NET is not supported on Media Center or Tablet editions Installer will optionally install NET Framework Visit www co2meter com pages downloads to download our complimentary GasLab software to your computer You can also download the GasLab user manual in PDF from this page Please read the GasLab user manual ...

Page 8: ...rs for the sensor automatically Software Capabilities Our GasLab software will allow users to Manage and download logs Configure sensor Adjust logging intervals Calibrate the sensor Collect data real time Powering the Metter This meter can be powered by four 4 AA size batteries not included but sampling units can also be powered by the included 6VDC international wall power supply These units shou...

Page 9: ...ved at a later time for analysis When combined with our GasLab software it allows you to see data in real time on your computer s screen Its built in 1 000 hour 10 000 hour optional sampling micro pump ensures reliable operation for persistent readings based on user selected time intervals Among others this meter is ideal for the following applications Incubation Fermentation monitoring and valida...

Page 10: ...Oxygen Sampling Data Logger User s Manual Page 10 of 24 Instrument Overview ...

Page 11: ...der to be as productive as possible in a short period of time Please read the Important Safeguards on page 4 before continuing IMPORTANT Follow these instructions to ensure proper set up 1 Download the GasLab software to your computer as shown in the GasLab Software section on page 7 2 Power the meter using the included universal power supply Turn ON the meter 3 Wait until the display changes colo...

Page 12: ...ws the labeled components of these meters as referenced throughout the rest of this manual Insert four 4 AA size batteries or connect to the included power supply This meter has an internal coin cell CR 2032 3V battery backup for the real time clock This battery is inserted in the factory for your convenience and should last the lifetime of the product In order to initialize data logging the unit ...

Page 13: ...nt information before making changes to the configuration of any of these ports You can reach us during regular business hours at 386 256 4910 or send us a message to support co2meter com Time Date Settings Here you can change the date and time on the meter for more accurate reference on the records The default time of your meter could vary depending on your location Review these settings before y...

Page 14: ...calibrate 3 Set the logging interval as desired using GasLab 4 Turn the Log switch to the ON position in order to start the logging session 5 When data logging is completed turn the Log switch to the OFF position in order to stop the logging session and disconnect the meter NOTE Refer to the GasLab User s Manual for more information Logging Settings Here you can change the logging interval and dur...

Page 15: ...p with Environmental Exhaust Figure 5 Opened loop operation We recommend installing the included humidity contaminate filter to ensure the sensing chamber and pump baffle stays clear and corrosion free Figure 6 Filter Installation and Orientation SAMPLING DATALOGGER ...

Page 16: ... set of AA batteries Duracell brand recommended or connect the included power supply 2 Install filters and tubing included as shown on Figure 6 3 Turn the meter ON by pressing the power switch once 4 For data logging models turn logging ON by pressing the LOG switch once Once the sampling process is completed data can be downloaded to your computer using our GasLab software ...

Page 17: ...f Calibration A Certificate of Calibration is required for products in ISO 9000 FDA regulated government contracted or other industries that requires traceability To order factory calibration you must first get a Return Material Authorization RMA number This number insures we know your unit is coming back to CO2Meter and returned to you the same day if possible To obtain an RMA number call 877 256...

Page 18: ...tificate at the same time you place your order https www co2meter com products certificate calibration Calibration Procedure The CM 0505 is calibrated using 100 Nitrogen Gas and Ambient Air 1 Install AA batteries and verify CM 0505 operates on battery power 2 Connect CM 0505 to 6V power supply and verify operation 3 Connect data cable from USB port on the computer to CM 0505 4 Turn ON CM 0505 and ...

Page 19: ...e Serial Number into the Name box 24 Use Snipping tool to record calibration graphic 25 Select Configure Sensor button 26 Select Data Logs tab 27 Verify Logging Interval is 5 28 Verify Device Time is correct 29 Select Clear Logs button 30 In the Clear Logs window select Yes 31 Verify Erasing Log on CM 0505 display 32 Close the Configure AQ500 window 33 Turn OFF Logging Switch on CM 0505 34 Turn OF...

Page 20: ... 22 to 140 F Life Expectancy 5 years Maintenance Interval No maintenance required Data Logging Data Points 60 000 Interval Period From every 2 seconds to every 18 hours Data Recorded Date time O2 Pump Characteristics Life Expectancy 1 000 hours basic pump Max Flow STP 0 40 liters min Max Vacuum 150 mbar Max Pressure 500 mbar Max System Pressure 1 atm 1 bar Power Batteries Four 4 AA sized Alkaline ...

Page 21: ...ain The display is blue or showing error The meter s EEPROM has been corrupted Improper power supply Please email our technical support for further assistance The GasLab software doesn t start 1 Your software might be out of date Update your software online at our website or by clicking on the CHECK FOR UPDATES under the Help menu Make sure your computer meets the minimum system requirements Slow ...

Page 22: ...Control Panel For Windows 8 press the Windows key and x at the same time to bring up the start menu then click on Device Manager For Windows 7 open the start menu and type Device Manager in the search bar 5 Find the Unrecognized USB Device in the list it usually but not always has a yellow triangle icon 6 Right click the Unrecognized USB Device item and select Update Driver Software 7 Select Brows...

Page 23: ... advantage of this warranty the product must be returned to CO2Meter com at your expense If after examination we determine the product is defective we will repair or replace it at no additional cost to you This warranty does not cover any products that have been subjected to misuse neglect accident modifications or repairs by you or by a third party No employee or reseller of CO2Meter com s produc...

Page 24: ...etermine that the product failed because of improper use water damage dropping tampering electrical damage etc or if it is beyond the warranty date we will inform you of the cost to fix or replace the product Contact Us We are here to help For information or technical support please contact us support co2meter com 386 256 4910 Technical Support 386 872 7665 Sales www co2meter com Address CO2Meter ...

Reviews: