5.2 Changing the Recording Settings
79
Chapte
r 5
Changing
Se
ttings (as
neces
sary)
5
Sets the TIME PLOT interval (recording interval).
"Key operation during configuration" (p.77)
The time series graph recording time varies with the recorded parameters and TIME PLOT interval setting.
Recording times (reference value) for a Z4001 SD Memory Card 2 GB
(Repeat Record: 1 week, Repeat Number: 55 times)
• Figures indicate the amount of recording time after the SD memory card has been initialized.
• Recording times do not account for event data and screen copy data. Recording times may be short-
ened when event data and screen copy data are stored on the card.
• Recording times are not dependent on connections.
• When repeated recording is set to
[OFF]
, the maximum recording time is 35 days.
• When repeated recording is set to
[1 Day]
, the maximum recording time is 366 days.
• When repeated recording is set to
[1 Week]
, the maximum recording time is 55 weeks.
• Harmonics order data is not saved for
[Power]
, but it is saved in THD.
Outputs the display image to the SD memory card at the
set screen copy interval.
"Key operation during configuration" (p.77)
TIME PLOT Interval
Setting Contents:(
∗
: Default setting)
1/ 3/ 15/ 30 second(s), 1
∗
/ 5 /10/ 15/ 30 minute(s) ,
1/2 hour(s), 150/180/1200 cycle
The 150 cycle (50 Hz) and 180 cycle (60 Hz) settings provide the TIME PLOT intervals
required for IEC 61000-4-30-compliant measurement. You can select 150 cycles (mea-
surement frequency of 50 Hz), 180 cycles (60 Hz), or 1200 cycles (400 Hz).
TIME PLOT interval
Recording parameter setting
All data
(Saves all data)
Power and Harmonic
(Saves RMS values and
harmonics)
Power
(Saves RMS values only)
1 second
16.7 hours
23.4 hours
13.2 days
3 seconds
2.1 days
2.9 days
39.7 days
15 seconds
10.4 days
14.6 days
198.4 days
30 seconds
20.9 days
29.3 days
55 weeks
1 minute
41.7 days
58.6 days
55 weeks
5 minutes
208.6 days
292.8 days
55 weeks
10 minutes
55 weeks
55 weeks
55 weeks
15 minutes
55 weeks
55 weeks
55 weeks
30 minutes
55 weeks
55 weeks
55 weeks
1 hour
55 weeks
55 weeks
55 weeks
2 hours
55 weeks
55 weeks
55 weeks
150/180 /1200 cycle (Approx. 3 sec) 2.1 days
2.9 days
39.7 days
When the memory is full
The PQ3198 stops recording data to the SD memory card.
Measuring for an extended period of time.
If repeated recording is enabled and the recording count set, the instrument can make measure-
ments for up to 55 weeks.
See:
Long-term measurements over 1 month or longer: Enable repeated recording (see "Repeat Record" (p.81)).
Screen copy interval
Setting Contents:(
∗
: Default setting)
OFF
∗
/5/ 10/ 30 minute(s)/ 1/ 2 hour(s)
HIOKI PQ3198A961-03
Summary of Contents for PQ3198
Page 2: ...HIOKI PQ3198A961 03...
Page 6: ...Contents iv HIOKI PQ3198A961 03...
Page 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Page 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Page 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Page 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Page 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Page 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Page 296: ...A32 HIOKI PQ3198A961 03...
Page 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Page 300: ...Index Index 4 HIOKI PQ3198A961 03...
Page 301: ...HIOKI PQ3198A961 03...
Page 302: ...HIOKI PQ3198A961 03...
Page 303: ...HIOKI PQ3198A961 03...
Page 304: ...www hioki com HIOKI PQ3198A961 03...