13.2 Input Specifications/Output Specifications/Measurement Specifications
200
Measurement accuracy
Same as for RMS voltage refreshed each half-cycle
Period: Within half a cycle of the start accuracy time, within half a cycle of the end accuracy time
(not specified for 400 Hz measurement)
Event threshold
Percentage of the nominal voltage or percentage of the slide reference voltage (user-selectable)
Event IN
Start of the waveform for which the RMS voltage refreshed each half-cycle exceeded the threshold
in the positive direction
Event OUT
Start of the waveform for which the RMS voltage refreshed each half-cycle fallen below the value
calculated by subtracting the hysteresis from the threshold
Multiple-phase system treatment
Starts when any of the U1 to U3 channels experiences a swell and ends when the swell has ended
for all channels.
Saved waveforms
Event waveforms
Fluctuation data
Saves the data of RMS voltage refreshed each half-cycle and RMS current refreshed each half-
cycle obtained in the period between 0.5 sec prior to the event IN and 29.5 sec after the event IN.
When set to 400 Hz, saves the data of RMS voltage refreshed each half-cycle and RMS current
refreshed each half-cycle obtained in the period between 0.125 sec prior to the event IN and 7.375
sec after the event IN.
(6) Swell (Swell)
(7) Dip (Dip)
Measurement method
Compliant with IEC61000-4-30
During 50 Hz/60 Hz measurement, a dip is detected when the RMS voltage refreshed each half-
cycle falls below the threshold.
During 400 Hz measurement, a dip is detected when the minimum of 4 RMS voltage values occur-
ring within 10 ms (values calculated for one 400 Hz waveform) falls below the threshold.
Displayed item
Dip depth : Worst value for RMS voltage refreshed each half-cycle [V]
Dip duration: Period from the time a U1 to U3 dip is detected until the reading exceeds the value
obtained by subtracting the hysteresis from the threshold in the positive direction
Measurement range
600.00 V
Measurement accuracy
Same as for RMS voltage refreshed each half-cycle
Period: Within half a cycle of the start accuracy time, within half a cycle of the end accuracy time
(not specified for 400 Hz measurement)
Event threshold
Percentage of the nominal voltage or percentage of the slide reference voltage (user-selectable)
Event IN
Start of the waveform for which the RMS voltage refreshed each half-cycle fallen below the thresh-
old
Event OUT
Start of the waveform for which the RMS voltage refreshed each half-cycle fallen below the value
calculated by adding the hysteresis to the threshold.
Multiple-phase system treatment
Starts when any of the U1 to U3 channels experiences a dip and ends when the dip has ended for
all channels.
Saved waveforms
Event waveforms
Fluctuation data
RMS data refreshed each cycle is saved from 0.5 s before to 29.5 s after the EVENT IN. When set
to 400 Hz, RMS data refreshed each cycle is saved from 0.125 s before to 7.375 s after.
(8) Interruption (Intrpt)
Measurement method
Compliant with IEC61000-4-30
During 50 Hz/60 Hz measurement, an interruption is detected when the RMS voltage refreshed
each half-cycle falls below the threshold.
During 400 Hz measurement, an interruption is detected when the minimum of 4 RMS voltage val-
ues occurring within 10 ms (values calculated for one 400 Hz waveform) exceeds the threshold in
the negative direction.
Displayed item
Interruption depth
: Worst value for RMS voltage refreshed each half-cycle [V]
Interruption duration : Period from the time a U1 to U3 interruption is detected until the reading ex-
ceeds the value obtained by adding the hysteresis to the threshold
Measurement range
600.00 V
HIOKI PQ3198A961-03
Summary of Contents for PQ3198
Page 2: ...HIOKI PQ3198A961 03...
Page 6: ...Contents iv HIOKI PQ3198A961 03...
Page 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Page 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Page 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Page 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Page 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Page 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Page 296: ...A32 HIOKI PQ3198A961 03...
Page 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Page 300: ...Index Index 4 HIOKI PQ3198A961 03...
Page 301: ...HIOKI PQ3198A961 03...
Page 302: ...HIOKI PQ3198A961 03...
Page 303: ...HIOKI PQ3198A961 03...
Page 304: ...www hioki com HIOKI PQ3198A961 03...