13.2 Input Specifications/Output Specifications/Measurement Specifications
211
10
13
Cha
p
ter 13 Spe
cificatio
ns
Measurement accuracy with a fundamental wave of 50 Hz/60 Hz
Harmonic input
Measurement accuracy
Notes
Voltage
1% or greater of nominal voltage
The 0th order:
±
0.3% rdg.
±
0.08% f.s.
1
st
or higher :
±
5.00% rdg.
Defined for a nominal voltage of
100 V or greater.
<1% of nominal voltage
The 0th order:
±
0.3% rdg.
±
0.08% f.s.
1
st
or higher :
±
0.05% of nominal voltage
Defined for a nominal voltage of
100 V or greater.
Current
The 0th order
:
±
0.5% rdg.
±
0.5% f.s.
1
st
to 20
th
:
±
0.5% rdg.
±
0.2% f.s.
21
st
to 50
th
:
±
1.0% rdg.
±
0.3% f.s.
Add current sensor accuracy.
Power
The 0th order
:
±
0.5% rdg.
±
0.5% f.s.
1
st
to 20
th
:
±
0.5% rdg.
±
0.2% f.s.
21
st
to 30
th
:
±
1.0% rdg.
±
0.3% f.s.
31
st
to 40
th
:
±
2.0% rdg.
±
0.3% f.s.
41
st
to 50
th
:
±
3.0% rdg.
±
0.3% f.s.
Add current sensor accuracy.
Measurement accuracy with a fundamental wave of 400 Hz
Harmonic input
Measurement accuracy
Notes
Voltage
The 0th order
:
±
0.5% rdg.
±
0.08% f.s.
1
st
to 2
nd
:
±
0.5% rdg.
±
0.20% f.s.
3
rd
to 6
th
:
±
1.0% rdg.
±
0.30% f.s.
7
th
to 10
th
:
±
5.0% rdg.
±
0.30% f.s.
Current
The 0th order:
±
0.5% rdg.
±
0.5% f.s.
1
st
to 2
nd
:
±
0.5% rdg.
±
0.2% f.s.
3
rd
to 6
th
:
±
1.0% rdg.
±
0.3% f.s.
7
th
to 10
th
:
±
5.0% rdg.
±
0.3% f.s.
Add current sensor accuracy.
Power
The 0th order:
±
0.5% rdg.
±
0.5%f.s.
1
st
to 2
nd
:
±
0.5% rdg.
±
0.2% f.s.
3
rd
to 6
th
:
±
1.0% rdg.
±
0.3% f.s.
7
th
to 10
th
:
±
7.0% rdg.
±
0.3% f.s.
Add current sensor accuracy.
(29) Inter-harmonic voltage and inter-harmonic current (Uiharm, Iiharm)
Measurement method
Compliant with IEC61000-4-7:2009
After harmonic analysis, harmonic voltage and harmonic current are displayed by adding as inter-
harmonic contents with the harmonic contents according to harmonic order
Measurement accuracy is defined for input that is 10% to 200% of IEC61000-2-4 Class 3.
Analysis window width
10 cycles (50 Hz) or 12 cycles (60 Hz)
Number of window points
Rectangular, 4096 points
Displayed item
From the 0.5th to 49.5th order (with a fundamental wave of 40 Hz to 70 Hz)
Select either RMS or content percentage (When using content percentage, Zero-display range
causes all orders to be given as 0% when the RMS value is 0.)
Measurement range
Inter-harmonic voltage: U1 to U4, 600.00 V
Inter-harmonic current: I1 to I4, Varies with used current sensor (see input specifications).
Measurement accuracy
Inter-harmonic voltage (Defined for a nominal voltage of at least 100 V.)
1% or greater of harmonic input nominal voltage::
±
5.00% rdg.
<1% of harmonic input nominal voltage:
±
0.05% of nominal voltage
Inter-harmonic current: Unspecified
Event threshold
N/A
Constraints
Not displayed for 400 Hz measurement.
HIOKI PQ3198A961-03
Summary of Contents for PQ3198
Page 2: ...HIOKI PQ3198A961 03...
Page 6: ...Contents iv HIOKI PQ3198A961 03...
Page 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Page 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Page 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Page 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Page 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Page 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Page 296: ...A32 HIOKI PQ3198A961 03...
Page 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Page 300: ...Index Index 4 HIOKI PQ3198A961 03...
Page 301: ...HIOKI PQ3198A961 03...
Page 302: ...HIOKI PQ3198A961 03...
Page 303: ...HIOKI PQ3198A961 03...
Page 304: ...www hioki com HIOKI PQ3198A961 03...