13.2 Input Specifications/Output Specifications/Measurement Specifications
203
10
13
Cha
p
ter 13 Spe
cificatio
ns
(14) RMS voltage (Urms)
Measurement method
AC+DC True RMS type
IEC61000-4-30 compliant: 10 cycles (50 Hz) or 12 cycles (60 Hz) (approx. 200 ms)
During 400 Hz measurement, calculated from 80 cycles (approx. 200 ms)
When set to 3P3W3M/3P4W/3P4W2.5E, the phase voltage/line voltage setting is applied to the
RMS voltage Urms.
Includes Zero-display range.
Displayed item
RMS voltage for each channel and AVG (average) RMS voltage for multiple channels
(for more information,see "13.8 Calculation Formula" (p.232))
Measurement range
600.00 V
Measurement accuracy
When the measurement frequency is set to 50 Hz/60 Hz
With input of 10 V to 660 V: ±0.1% of the nominal voltage; defined for a nominal input voltage
(Udin) of 100 V or greater.
With input outside the range of 10 V to 660 V or a nominal input voltage (Udin) of less than 100 V:
±0.2% rdg. ±0.08% f.s.
When the measurement frequency is set to 400 Hz
±0.2% rdg.±0.16% f.s.
Event threshold
The upper and lower limits can be separately set within the range of 0 to 780 V (lower limit < upper
limit) (value before setting VT ratio)
When set to 3P3W3M/3P4W/3P4W2.5E, the phase voltage/line voltage setting is applied.
Sense
Set from 0 V to 600 V.
Event IN
Start of the approx. 200 ms aggregation during which the reading exceeded the upper limit or fallen
below the lower limit
Event OUT
Start of the approx. 200 ms aggregation during which the reading was less than upper limit minus
hysteresis after being greater than the upper limit or was greater than lower limit plus hysteresis
after being less than the lower limit
Multiple-phase system treatment
Separate by channel
Saved waveforms
Event waveforms
(15) Voltage DC value (Udc)
Measurement method
Average value during approx. 200 ms aggregation synchronized with the reference channel (CH4 only)
Includes Zero-display range.
Displayed item
Voltage DC value
Measurement range
600.00 V
Measurement accuracy
±
0.3% rdg.
±
0.08% f.s.
Event threshold
0 V to 1200 V
The difference between the positive and negative waveform peak values in the 200 ms aggrega-
tion is compared to the threshold to generate DC fluctuation events.
Event IN
Start of the 200 ms aggregation in which the threshold was exceeded
Event OUT
Start of the first 200 ms aggregation after the IN state in which the threshold was not exceed
Multiple-phase system treatment
None
Saved waveforms
Event waveforms
HIOKI PQ3198A961-03
Summary of Contents for PQ3198
Page 2: ...HIOKI PQ3198A961 03...
Page 6: ...Contents iv HIOKI PQ3198A961 03...
Page 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Page 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Page 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Page 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Page 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Page 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Page 296: ...A32 HIOKI PQ3198A961 03...
Page 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Page 300: ...Index Index 4 HIOKI PQ3198A961 03...
Page 301: ...HIOKI PQ3198A961 03...
Page 302: ...HIOKI PQ3198A961 03...
Page 303: ...HIOKI PQ3198A961 03...
Page 304: ...www hioki com HIOKI PQ3198A961 03...