13.2 Input Specifications/Output Specifications/Measurement Specifications
201
10
13
Cha
p
ter 13 Spe
cificatio
ns
Measurement accuracy
Same as for RMS voltage refreshed each half-cycle
Period: Within half a cycle of the start accuracy time, within half a cycle of the end accuracy time
(not specified for 400 Hz measurement)
Event threshold
Percentage of the nominal voltage
Event IN
Start of the waveform for which the RMS voltage refreshed every cycle exceeded the threshold in
the negative direction
Event OUT
Start of the waveform for which the RMS voltage refreshed each half-cycle exceeded the value
calculated by adding the hysteresis to the threshold
Multiple-phase system treatment
Starts when all of the U1 to U3 channels experience an interruption and ends when the interruption
ends for any of the channels.
Saved waveforms
Event waveforms
Fluctuation data
RMS data refreshed each cycle is saved from 0.5 s before to 29.5 s after the EVENT IN. When set
to 400 Hz, RMS data refreshed each cycle is saved from 0.125 s before to 7.375 s after.
(8) Interruption (Intrpt)
(9) Instantaneous flicker value (Pinst)
Measurement method
As per IEC61000-4-15
User-selectable from 230 V lamp/120 V lamp (when Pst and Plt are selected for flicker measure-
ment)
Displayed item
Instantaneous flicker value
Measurement range, resolution
99.999, 0.001
Measurement accuracy
-
Event threshold
N/A
(10) Frequency 200 ms (Freq)
Measurement method
Reciprocal method
Calculated as the reciprocal of the accumulated whole-cycle time during approx. 200 ms period of
10, 12, or 80 U1 (reference channel) cycles.
Displayed item
Frequency 200 ms
Measurement range
When the measurement frequency is set to 50 Hz/60 Hz: 70.000 Hz
When the measurement frequency is set to 400 Hz
: 440.00 Hz
Measurement accuracy
When the measurement frequency is set to 50 Hz/60 Hz:
±
0.020 Hz or less
When the measurement frequency is set to 400 Hz
:
±
0.20 Hz or less (with input voltage of
4% f.s. to 110% f.s.)
Event threshold
Specified as deviation from 0.1 Hz to 9.9 Hz in 0.1 Hz increments
Event IN
Start of approx. 200 ms aggregation in which
±
threshold was exceeded
Event OUT
Start of approx. 200 ms aggregation in which reading returned to
±
(threshold - 0.1 Hz)
Note: Equivalent to 0.1 Hz frequency hysteresis.
Multiple-phase system treatment
None
Saved waveforms
Event waveforms
(11) 10-sec frequency (Freq10s)
Measurement method
Reciprocal method
Calculated as the reciprocal of the accumulated whole-cycle time during the specified 10 s period
for U1 (reference channel) as per IEC61000-4-30.
Displayed item
10-sec frequency
HIOKI PQ3198A961-03
Summary of Contents for PQ3198
Page 2: ...HIOKI PQ3198A961 03...
Page 6: ...Contents iv HIOKI PQ3198A961 03...
Page 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Page 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Page 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Page 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Page 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Page 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Page 296: ...A32 HIOKI PQ3198A961 03...
Page 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Page 300: ...Index Index 4 HIOKI PQ3198A961 03...
Page 301: ...HIOKI PQ3198A961 03...
Page 302: ...HIOKI PQ3198A961 03...
Page 303: ...HIOKI PQ3198A961 03...
Page 304: ...www hioki com HIOKI PQ3198A961 03...