A
10
Inrush current
Measurement method:
• Events will be detected when the current RMS Inrush exceeds the threshold value.
• For 400 Hz measurement, events are detected when the maximum of 4 RMS current values existing within the same
10 ms period (calculated values for one 400 Hz waveform) is greater than the threshold in the positive direction.
Event IN and OUT:
Event IN
: Time of start of channel half-cycle voltage waveform in which the RMS current refreshed each half-cycle
was greater than the threshold.
Event OUT : Time at the start of the voltage half-cycle waveform in which the RMS current refreshed each half-cycle ex-
ceeded (threshold - hysteresis) in the negative direction
Thresh-
old
Hysteresis
Inrush
EVENT
IN
EVENT
OUT
0.5 s
29.5 s
The current RMS Inrush within the period between 0.5 s befrore and 29.5 s after the event, as fluctu-
ating data, is saved.
Event IN
Event OUT
Current waveform
Voltage waveform
HIOKI PQ3198A961-03
Summary of Contents for PQ3198
Page 2: ...HIOKI PQ3198A961 03...
Page 6: ...Contents iv HIOKI PQ3198A961 03...
Page 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Page 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Page 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Page 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Page 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Page 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Page 296: ...A32 HIOKI PQ3198A961 03...
Page 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Page 300: ...Index Index 4 HIOKI PQ3198A961 03...
Page 301: ...HIOKI PQ3198A961 03...
Page 302: ...HIOKI PQ3198A961 03...
Page 303: ...HIOKI PQ3198A961 03...
Page 304: ...www hioki com HIOKI PQ3198A961 03...