8.4 Analyzing Transient Waveforms
149
Chapter 8 Checking Events (EVENT screen)
8
• The transient waveform consisting the fundamental component 50 Hz/60 Hz from a waveform sampled
at 2 MHz.
• The transient value measured from the waveform obtained by eliminating the fundamental component
50 Hz/60 Hz from having a sampled waveform pass through a highpass filter (fc = 5 kHz).
• Since voltage waveform display data reduced to 20 kS/s, the effect from a transient waveform may not
be reflected to the voltage waveforms.
8.4 Analyzing Transient Waveforms
Displaying transients
Accept setting
The display will switch to the
[VIEW]
screen, and the
waveform at the time of the
event will be displayed.
Select an event for
which “
Tran
” is
shown on the
event list or event
details list
Return to event list
[Event]
screen
[Event]
[List]
Year, month, day; time;Rising
transient voltage value ; and
duration of transient
[Transient]
Voltage/Transient waveforms display
Voltage
Waveform
Transient waveform
To enlarge or reduce the transient waveform (p.150)
To scroll transient waveform data (p.151)
Select with the
F
key.
HIOKI PQ3198A961-03
Summary of Contents for PQ3198
Page 2: ...HIOKI PQ3198A961 03...
Page 6: ...Contents iv HIOKI PQ3198A961 03...
Page 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Page 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Page 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Page 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Page 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Page 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Page 296: ...A32 HIOKI PQ3198A961 03...
Page 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Page 300: ...Index Index 4 HIOKI PQ3198A961 03...
Page 301: ...HIOKI PQ3198A961 03...
Page 302: ...HIOKI PQ3198A961 03...
Page 303: ...HIOKI PQ3198A961 03...
Page 304: ...www hioki com HIOKI PQ3198A961 03...