13.7 Other Functions
231
10
13
Cha
p
ter 13 Spe
cificatio
ns
13.7 Other Functions
-1. Warning functions
Wiring check
Checks connections and current sensor reverse connections as well as phase order on the con-
nection diagram screen.
Out of range
When the input exceeds the range by 130%, displays - - - - -.
Out of crest factor
When the waveform peak exceeds 2 times the voltage range or 4 times the current range, "crest
factor exceeded" is displayed.
Event check
Displays event icons when events occur.
Power supply status display,
Charge status display,
Battery strength display
See: "
Power supply status display" (p.30)
-2. Settings confirmation function
Function description
Press the
ESC
key during recording (including while in standby mode) to check the present set-
tings.
-3. Screen copy
Function description
Pressing the
COPY
key causes the instrument to store the screenshot displayed at the time onto
the SD memory card.
Data form
Compressed BMP format
File names
Auto generated, extension of “.bmp”
-4. Special key operation
Key lock function
Disables all key operation except for
POWER
switch and key lock cancelation.
Press and hold
ESC
key from at least 3 s to turn the function on and off.
If a passcode (in four digits or less) was entered at the time of the key lock activation, entering the
same passcode is required to disengage the key lock.
Display hold
Retains displayed values excluding the time
-5. Action in the event of an anomaly
Action in the event of a power
outage
With Model Z1003 Battery Pack sufficiently charged, the instrument starts to be powered by Model
Z1003, allowing continuous recording.
With Model Z1003 Battery Pack drained, the instrument stops recording. The settings used at the
time are backed up; however, integrated values are discarded.
When power is then recovered, the instrument starts recording and accumulating data anew in the
previous settings.
-6. Setup functionality
Function description
Sets the language when the instrument is turned on for the first time.
Boot key reset
Reverts all settings, including the language setting, to the factory defaults.
Turn on the instrument while holding down the
ENTER
and
ESC
keys.
HIOKI PQ3198A961-03
Summary of Contents for PQ3198
Page 2: ...HIOKI PQ3198A961 03...
Page 6: ...Contents iv HIOKI PQ3198A961 03...
Page 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Page 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Page 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Page 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Page 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Page 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Page 296: ...A32 HIOKI PQ3198A961 03...
Page 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Page 300: ...Index Index 4 HIOKI PQ3198A961 03...
Page 301: ...HIOKI PQ3198A961 03...
Page 302: ...HIOKI PQ3198A961 03...
Page 303: ...HIOKI PQ3198A961 03...
Page 304: ...www hioki com HIOKI PQ3198A961 03...