Method for verifying aggregation values required by IEC61000-4-30
IEC flicker
For IEC 61000-4-30 Plt values, use only the values shown with even numbered 2-hour intervals, and dis-
card the other Plt values. The other Plt values are provided for information only, and are not IEC 61000-
4-30 Plt values.
10-second frequency measurement
The parameter that is labeled "f10s" in TIMEPLOT is not an IEC 61000-4-30 measurement. User can find
frequency data for 10-second frequency measurement that is compliant with IEC 61000-4-30 here
(View>DMM>Voltage Freq10s)
IEC61000-4-30 Class A requires that regardless of the overall time interval, time clock accuracy must be
within ±20 ms for 50 Hz and within ±16.7 ms for 60 Hz. When accurate time synchronization using an
external signal is not possible, a tolerance of less than ±1 second over 24 hours is permitted, but regard-
less of the overall time interval, accuracy must be within ±20 ms for 50 Hz and ±16.7 ms for 60 Hz.
By synchronizing the PQ3198 with the PW9005 GPS Box, the instrument time can be synchronized to
UTC at a high degree of accuracy. In the event that accurate time synchronization using an external sig-
nal, such as that provided by the GPS unit, is not possible, the instrument incorporates a clock capable of
operating at a real-time accuracy of within ±1 second per day (within the specified operating temperature
and humidity range).
3-second aggregated values
(=150/180cycle data)
10-minute aggregated values
2-hour aggregated values
Magnitude of the
Supply Voltage
Applies to average value of
channel Urms values on the
[TIME PLOT]
-
[TREND]
screen.
Applies to average value of
channel Urms values on the
[TIME PLOT]
-
[TREND]
screen.
Applies to average value of
channel Urms values on the
[TIME PLOT]
-
[TREND]
screen.
Voltage
harmonics
Applies to average values on
the
[TIME PLOT]
-
[Harm-
Trend]
screen.
Applies to average values on
the
[TIME PLOT]
-
[Harm-
Trend]
screen.
Applies to average values on
the
[TIME PLOT]
-
[Harm-
Trend]
screen.
Voltage
inter-harmonics
Applies to average values for
each channel's orders on the
[TIME PLOT] - [Harm Trend]
-
[INTERHARM]
screen.
Applies to average values for
each channel's orders on the
[TIME PLOT] - [Harm Trend]
-
[INTERHARM]
screen.
Applies to average values for
each channel's orders on the
[TIME PLOT] - [Harm Trend]
-
[INTERHARM]
screen.
Supply Voltage
unbalance
Applies to average value of
unb0 and unb for Uunb on the
[TIME PLOT]
-
[TREND]
screen.
Applies to average value of
unb0 and unb for Uunb on the
[TIME PLOT]
-
[TREND]
screen.
Applies to average value of
unb0 and unb for Uunb on the
[TIME PLOT]
-
[TREND]
screen.
Measurement
conditions
• The TIME PLOT interval is
set to 150/180 cycles.
• During analysis, cursor
measurement is performed
after setting Tdiv to the min-
imum value.
• The order being checked for
harmonics and inter-har-
monics is selected and dis-
played.
• Recorded items for inter-
harmonics are set to
[All
data]
.
• Set the real-time control to
[Exactly]
.
• The TIME PLOT interval is
set to 10 minutes.
• During analysis, cursor
measurement is performed
after setting Tdiv to the min-
imum value.
• The order being checked for
harmonics and inter-har-
monics is selected and dis-
played.
• Recorded items for inter-
harmonics are set to
[All
data]
.
• Set the real-time control to
[Exactly]
.
• The TIME PLOT interval is
set to 2 hours.
• During analysis, cursor
measurement is performed
after setting Tdiv to the min-
imum value.
• The order being checked for
harmonics and inter-har-
monics is selected and dis-
played.
• Recorded items for inter-
harmonics are set to
[All
data]
.
• Set the real-time control to
[Exactly]
.
Time clock accuracy
HIOKI PQ3198A961-03
Summary of Contents for PQ3198
Page 2: ...HIOKI PQ3198A961 03...
Page 6: ...Contents iv HIOKI PQ3198A961 03...
Page 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Page 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Page 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Page 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Page 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Page 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Page 296: ...A32 HIOKI PQ3198A961 03...
Page 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Page 300: ...Index Index 4 HIOKI PQ3198A961 03...
Page 301: ...HIOKI PQ3198A961 03...
Page 302: ...HIOKI PQ3198A961 03...
Page 303: ...HIOKI PQ3198A961 03...
Page 304: ...www hioki com HIOKI PQ3198A961 03...