Basic operations |
iTernIS ADV
6.
8
By pressing the corresponding touch-screen button, the ventilator starts delivering
a flow of O
2
at 8 L/min through the Nebulizer exit, during the inspiratory phase. This
flow should be channeled through the patient circuit. Use the accessories included
in the nebulizer kit for this. The nebulizer kit is placed between the final inspiratory
segment of the patient circuit and the Y-shaped connection.
The flow provided by the Nebulizer is considered by the equipment to be an inspira
-
tory flow and is added to the flow coming from the patient outlet opening to complete
the programmed volumes (in VCV).
WARNING
• If the Nebulizer function is activated and the
flow provided by the Nebulizer outlet is NOT
directed to the patient circuit, volumes that
are lower than the adjusted values will be de
-
livered in VCV breaths.
Nebulization can be interrupted in three different ways:
• By time. 30 minutes after the function is activated.
• Voluntarily. By pressing the
[Nebulizer]
key.
• Automatically by the ventilator. If, during nebulization, the peak flow does not reach
an adequate value to be compatible with nebulization, the nebulization is interrupted.
This event is accompanied by low priority alarms (Interrupted Nebulization alarm).
Nebulization is enabled for all operative modes in ADL and PED patient categories.
For NEO-INF patients, only the TCPL and SIMV (TCPL) + PSV modes have an
option for nebulization with a continuous flow of 8 L/min.
Fig. 6-6
RESPIRATORY
MECHANICS Menu.
iTernIS ADV
offers the possibility of performing different tests to categorize the res-
piratory mechanics of the patient being ventilated. For further information, consult
the
Respiratory Mechanics
chapter, which describes all the tests in detail.
NOTE
• Keep in mind that some respiratory mechanics
tests may only be performed in certain operati-
ve modes and under certain conditions (see the
Respiratory Mechanics
chapter).
6.5.1.1 | Nebulizer
6.5.1.2 | Respiratory Mechanics
6.5.1 |
Direct Access
Summary of Contents for iTernIS ADV
Page 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Page 2: ......
Page 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Page 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Page 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Page 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Page 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Page 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Page 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Page 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Page 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Page 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Page 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Page 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Page 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Page 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Page 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Page 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Page 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Page 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Page 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Page 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Page 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Page 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Page 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Page 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Page 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Page 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Page 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Page 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Page 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Page 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Page 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Page 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Page 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Page 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Page 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Page 217: ......
Page 218: ......