Cleaning and maintenance |
iTernIS ADV
16.
6
NOTE
• Preventive maintenance is essential for the
structural and functional integrity of the ventila-
tor. Neglecting the adequate maintenance may
affect the quality of the mechanical ventilation
received by the patients.
• The ventilator lifespan has been established
in accordance with the performance of all the
scheduled maintenance services to be imple-
mented during this period (every 5000 hours of
usage or every year).
• Scheduled maintenance can only be executed by
HEYER or personnel authorized by HEYER Medi-
cal AG
Table 16-1
Suggested preventive maintenance
FREQUENCY
PART/ACCESSORY
PROCEDURE
Every 4 to 6 hrs.
Water traps of the patient
circuit.
Check for presence of
water in the traps. Drain
if necessary.
With each new patient.
Patient circuit, including
the expiratory set (valve +
expiratory pneumotacho-
graph).
Replace with a sterile cir-
cuit, and sterilize or discard
the outgoing circuit, or the
corresponding parts.
Daily.
Water filter for com
-
pressed air.
Check presence of water
in the filter. Drain if nec
-
essary.
Daily.
Patient circuit.
Check status of overall
circuit (hoses + accesso-
ries). Replace the parts
or accessories that are
not in optimal conditions,
and recalibrate the pa-
tient circuit.
Every 3 months.
Internal battery.
Disconnect the main power
cable. Verify that the icon
Full battery charge remains
for 10 minutes. If this does
not occur, contact your au-
thorized service.
Annually or every 5000
hrs. of usage (which-
ever comes first).
Complete ventilator.
Send the ventilator to an
authorized technical ser-
vice for scheduled main-
tenance of 5000 hrs.
Every 2 years or every
10000 hrs. of usage
(whichever comes
first).
Complete ventilator.
Send the ventilator to an
authorized technical ser-
vice for scheduled main-
tenance of 10000 hrs.
Every 3 years or every
15000 hrs. of usage
(whichever comes
first).
Complete ventilator.
Send the ventilator to an
authorized technical ser-
vice for scheduled main-
tenance of 15000 hrs.
It is very important that the air filter is kept in optimal working conditions. The com
-
pressed air quality affects the working order of the ventilator, as well as the patient’s
health.
16.3.2 |
Compressed air filter
16.3 | Maintenance
16.3.1 |
Suggested preventive
maintenance
Summary of Contents for iTernIS ADV
Page 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Page 2: ......
Page 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Page 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Page 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Page 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Page 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Page 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Page 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Page 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Page 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Page 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Page 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Page 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Page 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Page 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Page 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Page 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Page 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Page 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Page 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Page 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Page 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Page 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Page 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Page 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Page 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Page 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Page 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Page 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Page 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Page 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Page 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Page 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Page 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Page 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Page 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Page 217: ......
Page 218: ......