Cleaning and maintenance |
iTernIS ADV
16.
3
The ventilator
iTernIS ADV
and the parts associated to the patient circuit (hoses,
connectors, accessories, etc.) are delivered clean but not sterilized.
The parts which come into contact with the patient are easily disassembled for cle-
aning and subsequent sterilization, which should be done after every use.
PRECAUTION
• The ventilator is not autoclavable and is not
compatible with ethylene oxide. Do not subject
the equipment to any of these methods of steri-
lization.
NOTE
• The information provided to manage the cle
-
aning and sterilization of the different compo-
nents works as a guide. This can be adapted to
the protocols of each department or service, so
long as the warnings and cautions stated in this
chapter are respected.
• To discard all of the equipment, parts, or ele
-
ments provided by other suppliers, follow the
requirements of the institutional authority.
WARNING
• Do not clean the screen with harsh chemical
solvents, nor acid-based or alkaline subs
-
tances. Use a soft cloth soaked in a solution
of neutral detergent and water. A solution
containing 80% of isopropyl alcohol can
also be used. Never use substances which
contain ammonia.
• Do not spray the cleaning product directly
onto the screen. Apply the chosen substan
-
ce onto a soft cloth with which to wipe the
screen.
All parts of the patient circuit must be in perfect aseptic conditions at time of use.
All items delivered by HEYER Medical AG. with the ventilator, which are not pertai-
ning to the same manufacturing, must be cleaned and sterilized according to the
specifications of their manufacturers. Read and adhere to these instructions, in
order to always use the items duly equipped with the ventilator
iTernIS ADV
.
For cleaning prior to sterilization, the patient circuit must be completely disassem-
bled in order to expose all surfaces. It is inadmissible that remnants of organic
material be left stuck onto any segment of the circuit. See Fig. 16-1 showing a
diagram of the patient circuit.
16.1 | Cleaning and
Sterilization
16.1.1 |
General instructions
16.1.2 |
Touch screen
16.1.3 |
Patient circuit
Summary of Contents for iTernIS ADV
Page 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Page 2: ......
Page 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Page 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Page 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Page 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Page 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Page 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Page 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Page 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Page 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Page 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Page 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Page 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Page 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Page 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Page 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Page 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Page 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Page 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Page 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Page 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Page 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Page 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Page 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Page 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Page 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Page 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Page 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Page 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Page 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Page 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Page 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Page 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Page 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Page 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Page 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Page 217: ......
Page 218: ......