![Heyer iTernIS ADV Operator'S Manual Download Page 198](http://html1.mh-extra.com/html/heyer/iternis-adv/iternis-adv_operators-manual_2128282198.webp)
Technical specifications|
iTernIS ADV
17.
8
17.11.2 |
Manufacturer
Statement:
Electromagnetic Immunity
Table 17-11
Electromagnetic Emissions Statement
iTernIS ADV
is intended to be used in the specified electromagnetic environ
-
ment. The customer and/or user of
iTernIS ADV
must ensure it is used in an
electromagnetic environment as stated in this Table.
Emission Test
Compliance
Electromagnetic Environment
RF CISPR 11
Emissions
Group 1
iTernIS ADV
uses RF energy only for its
internal operations. Therefore, its RF emis-
sions are very low and they are unlikely to
cause any interference in nearby electronic
equipment.
RF CISPR 11
Emissions
Class A
iTernIS ADV
can be used in all establish-
ments, except in households, and those that
are connected directly to the public low vol-
tage power supply network feeding buildings
used for domestic purposes.
IEC 61000-3-2
harmonic emis-
sions
Class A
IEC 61000-
3-3 Voltage
fluctuations /
discontinuous
emissions
According to
specifications
Table 17-12
Electromagnetic immunity Statement
iTernIS ADV
is intended to be used in the electromagnetic specified environ
-
ment. The customer and/or user of
iTernIS ADV
must ensure it is used in an
electromagnetic environment as stated in this Table.
Immunity test
IEC 60601-1-2
Test level
Compliance
level
Electromagnetic
environment
IEC 610000-4-
4 Electrosta-
tic Discharge
(ESD)
Contact: ± 6 kV.
Air: ± 8 kV.
Contact: ± 6 kV.
Air: ± 8 kV.
Floors must be
wood, concrete, or
ceramic. If floors are
covered with synthetic
material, the relative
humidity must be at
least 30%.
IEC 61000-4-4
Fast electrical
transient in
bursts
± 2 kV for power
lines.
± 1 kV for input/
output lines.
± 2 kV for power
lines.
± 1 kV for input/
output lines.
The quality of the
electrical network must
be equivalent to that of
a typical commercial or
hospital environment.
IEC 61000-4-5
Shock waves
Differential mode:
±1 kV.
Shared mode: ±
2 kV.
Differential mode:
±1 kV.
Shared mode: ±
2 kV.
The quality of the
electrical network must
be equivalent to that of
a typical commercial or
hospital environment.
IEC 61000-
4-11 Voltage
dips, short
interruptions
and voltage
variations on
line power
input
< 5% UT (low >
95% of UT) for 0.5
cycles.
40% UT (low of
60% of UT) for 5
cycles.
70% UT (low of
30% of UT) for 25
cycles.
< 5% UT (low >
95% of UT) for 5 s.
< 5% UT (low >
95% of UT) for 0.5
cycles.
40% UT (low of
60% of UT) for 5
cycles.
70% UT (low of
30% of UT) for 25
cycles.
< 5% UT (low >
95% of UT) for 5 s.
The quality of the
electrical network must
be equivalent to that of
a typical commercial or
hospital environment.
If the user of
iTernIS
ADV
requires maintai-
ning operation during
power outages, we
recommend connecting
the system to an unin-
terrupted power supply
or a battery.
17.11.1 |
Manufacturer
Statement:
Electromagnetic Emissions
Summary of Contents for iTernIS ADV
Page 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Page 2: ......
Page 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Page 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Page 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Page 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Page 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Page 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Page 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Page 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Page 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Page 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Page 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Page 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Page 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Page 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Page 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Page 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Page 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Page 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Page 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Page 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Page 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Page 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Page 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Page 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Page 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Page 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Page 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Page 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Page 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Page 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Page 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Page 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Page 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Page 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Page 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Page 217: ......
Page 218: ......