Table A-1
Initial Calibration
Event
Possible Causes
Suggested Action
Leak smaller than 10
L/min
Faulty connections in the
patient circuit. Malfunctio-
ning elements of the cir-
cuit.
Check the integrity of the
patient circuit hoses and
accessories. Change the
damaged segment, if any.
Revise the overall status
of the expiratory set, es-
pecially the expiratory val-
ve diaphragm.
Leak greater than 10 L/
min
Y-shaped connector end
without adequate obstruc-
tion. Disconnected expi-
ratory pneumotachograph
hoses (P1 and P2). No
O
2
sensor. Faulty connec-
tions in the patient circuit.
Malfunctioning elements
of the circuit.
Verify the plug on the Y
connector. Check connec-
tions on P1 and P2. Verify
presence of the O
2
sensor.
Check the integrity of the
patient circuit hoses and
accessories. Replace the
damaged segment, if any.
Defective expiratory
set or airtightness loss
Poorly connected pneu-
motachograph hoses (P1
and P2). Inner membra-
ne damage of pneumota-
chograph. Partially obs-
tructed Y connector.
Check the cap on the Y
connector. Check connec-
tions on P1 and P2. Check
integrity of the pneumota-
chograph internal membra-
ne. Solve the failure and
execute a new calibration.
Failed calibration of O
2
sensor
Sensor exhausted, or
presence of incorrect gas
in the OXYGEN inlet.
Verify that the gas supply
is actually oxygen and
that its concentration is
adequate. Verify the sta-
tus of the O
2
sensor.
Oxygen sensor not
detected
The microprocessor does
not receive electrical sig-
nals from the sensor.
Check the sensor con-
nection. If the sensor
were properly connected,
this can signify that the
sensor is exhausted, and
if so, replace it.
ALARMS during cali-
bration
Low oxygen or air pres-
sure, or low pressure of
both gases. Power loss.
See Table for correspon-
ding alarm below.
Table A-2
High Priority Alarms
ALARM
Possible Causes
Suggested Action
Inop. Vent.
Microprocessor failure
Contact the authorized
technical service
Summary of Contents for iTernIS ADV
Page 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Page 2: ......
Page 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Page 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Page 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Page 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Page 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Page 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Page 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Page 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Page 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Page 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Page 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Page 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Page 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Page 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Page 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Page 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Page 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Page 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Page 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Page 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Page 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Page 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Page 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Page 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Page 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Page 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Page 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Page 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Page 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Page 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Page 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Page 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Page 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Page 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Page 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Page 217: ......
Page 218: ......