![Heyer iTernIS ADV Operator'S Manual Download Page 210](http://html1.mh-extra.com/html/heyer/iternis-adv/iternis-adv_operators-manual_2128282210.webp)
Appendices |
iTernIS ADV
A.
ii
Table A-5
Low Priority Alarms
ALARM
Possible causes
Suggested action
Maximum rate
Low alarm limit. Leaks
in the patient circuit.
Check the alarm limit, and correct
it if necessary. In the case of ma-
jor leaks, high frequency auto-
triggering can be generated with
the configured alarm limit. In this
case, identify the possible areas
of leakage and repair them.
PEEP loss
Incorrectly configured
alarm limit. Leaks.
Faulty expiratory val-
ve diaphragm.
Review the alarm limits and iden-
tify the possible areas of leaka-
ge. Lastly, check the status of the
valve diaphragm. If necessary to
replace the diaphragm, use an
authentic one.
Nebulization OFF
Insufficient peak flow,
operate with nebuliza-
tion.
The nebulization is reset if the
peak flow is enough to be com
-
patible with the nebulization.
Transporting
Lack of gas pressu-
re at the AIR inlet,
and activation of the
Transfer function. This
must be activated by
the user.
No corrective action is sugges-
ted. When the air supply is reset,
or when the user disables the
Transfer function, the alarm sig-
nals disappear.
Table A-6 Capnography Messages
Message
Possible Causes
Suggested Action
CO
2
no breathing
detected
Apnea condition Dis-
connected adapter of
the sensor, or of the
patient circuit. Dama-
ged or dirty adapter.
Check the capnograph adapter
connections. Clean if the win-
dows are very dirty.
CO
2
out of range
Measured value ex-
ceeds 150 mm Hg.
Perform a reset to zero.
Check the patient
adapter
Sensor out of the
adapter. Optical bloc-
king of the adapter
windows. Defective
reset to zero.
Check the presence of the sen-
sor in the adapter. Check that the
adapter windows are free of dirt.
In this case, clean the adapter,
allowing the sensor to release
any residual CO
2
from its interior,
and execute a new reset to zero.
Sensor failure
Possible disconnection.
Check the sensor electrical con-
nection with the ventilator. If the
error persists, contact the autho-
rized technical service.
Sensor not acti-
vated
Poorly executed com-
pensation of barome-
tric pressure and gases
during the initial calibra-
tion of the ventilator.
Activate sensor, or save para-
meters from the capnography
menu.
Zero required
Problems with the
reset to zero of the
capnograph. Dirt, for
example.
Repeat reset to zero.
Sensor overheated
Sensor exposed to ex-
ternal heat sources.
Check presence of heat sources
within proximity of the sensor,
such as lamps or stoves. If the
message persists, contact the
authorized technical service.
A.I.6 |
Capnography
Messages
A.I.5 |
Low Priority Alarms
Summary of Contents for iTernIS ADV
Page 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Page 2: ......
Page 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Page 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Page 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Page 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Page 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Page 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Page 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Page 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Page 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Page 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Page 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Page 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Page 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Page 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Page 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Page 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Page 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Page 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Page 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Page 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Page 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Page 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Page 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Page 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Page 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Page 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Page 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Page 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Page 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Page 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Page 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Page 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Page 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Page 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Page 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Page 217: ......
Page 218: ......