![Heyer iTernIS ADV Operator'S Manual Download Page 174](http://html1.mh-extra.com/html/heyer/iternis-adv/iternis-adv_operators-manual_2128282174.webp)
Capnography |
iTernIS ADV
14.
6
If the message ‘Sensor not activated’ is received, use this CAPNOGRAPHY menu
option. The ventilator will send barometric pressure and oxygen compensation va-
lues to the capnography system. This will initiate the sensor without the need for a
new activation.
In case the message ‘Zero required’ is received, a calibration of the sensor from
zero must be executed.
In the case of replacing the sensor adapter, it is recommended to perform a reset
to zero of the sensor.
To perform the sensor reset to zero, follow these steps:
• Disconnect the sensor from the patient circuit.
• Expose the assembly to ambient air (the adapter should not contain residues of
CO
2
while calibrating).
• Execute the following string:
[Menu]>>Tools>>Calibration>>Capnography>>
Sensor zero
.
• Check the message ‘Zero activated’ (on the CAPNOGRAPHY screen, the messa
-
ge ‘Zero in progress’ also appears). The procedure takes 30 s (maximum).
This is used in extraordinary situations when it is necessary to restart the sensor. This
function allows for the reset, without having to disconnect and reconnect the sensor.
Next, messages which may appear during the use of the capnography module are
defined, as well as detailed recommendations on necessary actions.
• ‘CO
2
NO BREATH DETECTED’
This message appears in the case of failing to detect breaths containing CO
2
.
• ‘CO
2
OUT OF RANGE’
Appears when the value of CO
2
is detected as being superior to that of 150 mmHg.
If this error persists, a reset to zero must be executed.
• ‘CHECK PATIENT AIRWAY ADAPTER’
This usually appears when the sensor is removed from the adapter or when there
is an optical blockage of those same windows. A flaw may also be caused by a
failure in the reset to zero (during the change of adapter or when there has been
CO
2
inside it at the time of reset to zero). Cleaning of the adapter is recommended,
or if it is clean, execute a new reset to zero.
• ‘SENSOR FAILURE’
This may appear when a deviation of the source current outside the factory cali-
bration occurs or if it is out of the performance specifications. Check if the sensor
is properly connected to the ventilator. If the error persists, contact the Authorized
Service.
14.4.3 |
Save sensor settings
14.4.4 |
Zero
14.4.5 |
Reset
14.5 | Messages
Summary of Contents for iTernIS ADV
Page 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Page 2: ......
Page 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Page 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Page 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Page 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Page 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Page 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Page 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Page 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Page 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Page 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Page 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Page 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Page 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Page 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Page 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Page 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Page 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Page 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Page 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Page 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Page 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Page 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Page 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Page 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Page 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Page 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Page 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Page 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Page 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Page 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Page 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Page 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Page 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Page 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Page 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Page 217: ......
Page 218: ......