Technical specifications |
iTernIS ADV
17.
5
Oxygen (con-
centration)
(%)
21 - 100
1*
50
Inspiratory
sensitivity
(Flow
= L/min; Pressure =
cmH
2
O)
Triggering by flow:
0.2 – 15
Triggering by pres-
sure: -0.5 a –20
By flow:
0.1/0.5/1
according to the
detection value.
By pressure: 0.5
ADL: Flow=3;
Pressure=-1.5
PED: Flow=3;
Pressure=-1.5
NEO-INF:
Flow=1; Pressu-
re=-1.5
Expiratory sensiti-
vity
(for modes with
PSV)
5% - 80% of the
peak flow
5%
25%
Programmable
inspiratory pau-
se (in VCV)
(s)
0 – 2
0,25
0 (NO)
Manual inspira-
tory pause
(s)
7 (maximum)
N/A
N/A
Manual expira-
tory pause
(s)
20 (maximum)
N/A
N/A
Sighs
(in VCV)
No. of sighs: 1/2/3
Rate: 5/10/15/20 per
hour
Added volume:
+0.1VT – +1.0VT
(The volume is
added to the confi
-
gured VT)
N/A
No. of sighs: 1.
Rate: 5 per hour
Added volume:
+30% (+0.3VT).
Activated: No
Inspiratory flow
waveform
Descending and
constant ramp (rec-
tangular)
N/A
Descending ramp
Leakage compen
-
sation in NIV
(L/
min)
Non adjustable
N/A
ADL: up to 50
PED: up to 30
Leakage com
-
pensation for
the rest of the
modes
(L/min)
Non adjustable.
N/A
ADL: up to 15
PED: up to 15
NEO-INF: up to
10
Tube compen-
sation
Tube: Endotracheal
or tracheostomy
Diameter: 4 – 12
mm
Compensate: 10%-
100%
N/A
Tube: Endotra-
cheal
Diameter: 8 mm
Compensate:
50%
Activated: NO
(1) An asterisk (*) next to some of the values in this column means that if you
press
[Ctrl]
before making a parameter change, larger or smaller increments
can be obtained according to the parameter in question.
(2) Initial factory values. By modifying mL/kg in setting IBW-based VT, these
values may change (see chapter Initial Verification and Calibration).
(3) These values are shown as a reference, and are to be found in VCV and
PCV modes. In order to know the initial values of each mode see the chapter
ADL/PED and NEO-INF Operative Modes.
(4) The Minute Volume is programmed only in MMV + PSV. For the rest of the
modes, the Minute Volume is the result of other parameters programming (see
maximum values by category in Table 17-5).
Summary of Contents for iTernIS ADV
Page 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Page 2: ......
Page 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Page 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Page 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Page 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Page 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Page 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Page 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Page 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Page 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Page 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Page 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Page 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Page 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Page 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Page 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Page 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Page 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Page 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Page 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Page 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Page 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Page 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Page 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Page 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Page 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Page 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Page 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Page 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Page 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Page 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Page 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Page 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Page 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Page 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Page 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Page 217: ......
Page 218: ......