![Heyer iTernIS ADV Operator'S Manual Download Page 181](http://html1.mh-extra.com/html/heyer/iternis-adv/iternis-adv_operators-manual_2128282181.webp)
Operational verification |
iTernIS ADV
15.
5
Table 15-2
Alarms Verification
TEST
ACTION
VERIFICATION
Disconnection
Disconnect the test lung from
the circuit.
The Disconnection alarm
signals should appear wi-
thout delay.
Auditory alarm
signal pause
During the active disconnec-
tion alarm, press the pause
key of the auditory alarm
signal.
The auditory alarm signal
should pause for a period
of 30 s.
Maximum
pressure
Regulate the alarm limit level
to 5 cmH
2
O above the current
pressure value. Then, during
an inspiratory phase, squeeze
the test lung.
The Maximum pressure
should appear immediately
and the peak pressure value
should not exceed that of the
established alarm limit.
Maximum VT
Regulate the VTmax alarm
limit below the reading of the
actual VT.
After 10 s the corresponding
alarm signals should appear.
Minimum VT
Regulate the VTmin alarm
limit to 20 mL. above the
current value.
After 10 s the corresponding
alarm signals should appear.
Maximum rate
Regulate this alarm limit to 20
rpm. Simulate spontaneous
breathing by squeezing and
releasing the test lung at a
rate greater than 20 times per
minute.
After 30 seconds of having
exceeded the 20 rpm, the
maximum rate alarm signals
should appear.
Apnea
Select the PSV ventilation
mode without changing the
menu Backup Ventilation.
Squeeze the test lung, only
once, to generate sponta-
neous breathing. Then sus-
pend the generated sponta-
neous breathing.
15 s after the last sponta-
neous breath, Apnea alarm
signals should appear, and
the ventilation mode must
change from PSV to the
configured mode of Backup
Ventilation.
Electric power
failure.
Disconnect the electric power
cable.
The corresponding alarm
signals should appear im-
mediately.
Maximum VM
(For ADL and
PED patient
category)
Select the MMV+PSV mode
and begin the ventilation
using the default parameters.
Regulate the VMmax alarm
limit to 4.0 L/min. Exceed this
limit by squeezing the test lung
repeatedly.
The corresponding alarm
signals should appear 10 s
after the ventilation starts to
exceed the limit.
Minimum VM
While in MMV +PSV, regulate
the VMmin alarm limit to 8.0 L/
min. Apply repeated pressure
to the test lung so as not to
exceed the alarm limit.
The corresponding alarm
signals should appear 10 s
after the ventilation starts to
go under the limit.
The tests which require user intervention have already been covered in the pre-
vious sections. Operative verification is an automatic control performed by the ven
-
tilator during the initial calibration. For further information on the nature of these
tests, refer to chapter
Start-up and Initial Calibration
.
15.5 | Operative Verification
Summary of Contents for iTernIS ADV
Page 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Page 2: ......
Page 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Page 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Page 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Page 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Page 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Page 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Page 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Page 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Page 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Page 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Page 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Page 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Page 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Page 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Page 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Page 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Page 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Page 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Page 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Page 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Page 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Page 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Page 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Page 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Page 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Page 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Page 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Page 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Page 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Page 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Page 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Page 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Page 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Page 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Page 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Page 217: ......
Page 218: ......