Respiratory mechanics |
iTernIS ADV
12.
11
[Respiratory Mechanics]>>PV flex.
After implementing this string, the test win-
dow appears on the screen. Pressing the rotary knob starts the maneuver for the
measurement. As can be seen in Fig. 12-6, there is a maximum pressure parame-
ter and a maximum volume parameter. These two values are the maximum allowed
during the procedure, and are used as a criterion for completing the delivery of the
flow. Whichever value is reached first, it will signal the end of the test. The configu
-
ration of these values is performed using the
[Menu]
key.
By pressing this key once, the Pmax limit changes color, indicating that it can be
changed by turning the rotary knob. It is accepted by pressing the knob, or the va-
lue is resumed by pressing
[Esc].
Press the
[Menu]
key again to change the Vmax
limit. The change is performed the same way as that described for Pmax.
Once the result is displayed on screen, it is possible to perform the test again.
During the implementation of the test, it is possible to cancel the maneuver by
pressing
[Esc].
Fig. 12-6
Picture of a PV flex test. The graph shown was
generated from a test mechanical lung with a compliance
change simulator. As a result, the changes in the curve can
be seen very clearly defined.
The Pimax is an index that evaluates the maximum contractile capacity of the ins-
piratory muscles, especially the diaphragm.
The Pimax is not only a reflection of the respiratory muscle function in isolation,
but it can also be affected by changes that occur anywhere in the genesis of mus-
cle contraction (central nervous system, conduction pathways, neuromuscular
junction, muscle mechanical situation, peripheral receptors, etc.).
This test is only enabled for the ADL and PED patient categories.
The Pimax can be performed with or without the cooperation of the patient. By
selecting
RESPIRATORY MECHANICS
in the test menu, the ventilator continues
in the selected mode and removes the PEEP.
The maneuver starts by analyzing the breathing cycle during two breaths, identi-
fying inspiration and expiration. During the last expiration, an inspiratory occlusion
12.6.3 |
Implementation
12.7 | Pi
max
12.7.1 |
Concepts
12.7.2 |
Procedure
Summary of Contents for iTernIS ADV
Page 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Page 2: ......
Page 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Page 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Page 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Page 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Page 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Page 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Page 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Page 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Page 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Page 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Page 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Page 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Page 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Page 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Page 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Page 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Page 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Page 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Page 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Page 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Page 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Page 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Page 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Page 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Page 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Page 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Page 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Page 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Page 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Page 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Page 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Page 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Page 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Page 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Page 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Page 217: ......
Page 218: ......