Operational verification |
iTernIS ADV
15.
4
Table 15-1
Functional verification
TEST
ACTION
VERIFICATION
Frequency
Monitor the amount of brea-
ths in one minute using the
clock second hand.
The value obtained from the
previous action must coincide
with the one monitored by the
ventilator.
Tidal volume (VT)
Contrast the VT value regu-
lated with the monitored one
for the preset value. Repeat
for a VT of 50% higher and
for a VT of 50% lower.
The monitored value must
be within ±10% of the regu-
lated value.
PEEP
Keep the preset value.
Verify that the base pressu-
re line remains stable.
Sensitivity
Apply pressure on the test
lung in order to generate an
inspiration.
The inspiration must be
generated.
Sensitivity
Regulate the sensitivity to
0.5 L/min.
The autocycling should not
be verified (allow for at least
five mandatory breaths).
Sensitivity
Regulate the sensitivity to
-0.5 cmH
2
O.
The autocycling should not
be verified (allow for at least
five mandatory breaths).
Manual inspira-
tion
Press
[Manual Insp].
The ventilator should initiate
a mandatory breath.
Flow waveform
Change the flow waveform.
On the flow chart, the profile
change of the flow wave
-
forms should appear.
Oxygen
Compare the monitored
O
2
concentration with the
regular concentration.
The monitored value should
be between 45% and 55%
Oxygen
Change the concentration
of O
2
up to a value of 100%.
The monitored oxygen con-
centration should be of at
least 95% after ten breaths.
NOTE
• During the VT test, the alarm limit of maximum
inspiratory pressure may be exceeded, and this
will activate the corresponding alarm signals. In
this case it is necessary to regulate the mini-
mum inspiratory pressure limit, so that the set
VT may be delivered correctly. Otherwise, this
alarm will limit the delivered VT, and the test will
not be executed.
The alarm verification tests are presented in Table 15-2. These tests are designed
to be carried out in the suggested order, and require that the value of a modified
ventilator parameter or alarm limit be restored to its original value once the test has
been concluded.
15.4 | Alarms Verification
Summary of Contents for iTernIS ADV
Page 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Page 2: ......
Page 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Page 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Page 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Page 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Page 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Page 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Page 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Page 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Page 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Page 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Page 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Page 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Page 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Page 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Page 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Page 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Page 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Page 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Page 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Page 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Page 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Page 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Page 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Page 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Page 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Page 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Page 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Page 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Page 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Page 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Page 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Page 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Page 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Page 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Page 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Page 217: ......
Page 218: ......