Error code
Description
Details and solution
8F
Ends of slider
2
movement not in expected
range
• Early firmware versions were not able to
use as much force to get or place media
in a magazine. Upgrading the library
firmware might resolve this issue.
• If the error consistently happens on the
same slot, try different tape cartridges in
that slot.
◦ If the failure remains with the same
slot, the magazine might be at fault.
◦ If the failure follows the tape
cartridge, the tape cartridge might be
at fault.
• If the device is performing an operation
that automatically returns a tape
cartridge to a certain slot, make sure that
another tape is not loaded in that slot.
• Run the wellness test
90
Ends of elevator
3
movement not in
expected range
• Remove all magazines and look for any
obstructions to the robot.
• Run the wellness test.
91
Ends of rotation
4
movement not in expected
range
• Remove all magazines and look for any
obstructions to the robot.
• Run the wellness test.
92
Ends of sled
5
movement not in expected
range
• Remove all magazines and look for any
obstructions to the robot.
• Run the wellness test.
93
Gripper
1
reached a position beyond
expected range
• Remove all magazines and look for any
obstructions to the robot.
• Run the wellness test.
94
Slider
2
reached a position beyond expected
range
95
Elevator
3
reached a position beyond
expected range
Table Continued