Assembly
81
1. Move the lift upward.
2.
Attach the protective shield (1) on the edge of the heating bath (2) and position so
that the (possibly available) cover can be easily opened.
3.
Turn the eccentric screw (3) by 180° and use it to attach the protective shield on the
heating bath.
When using the cover and the protective shield, take care that the two
parts do not jam.
1
2
Figure 4-29: Vapor temperature sensor
1
Plug
2
Seal
The vapor temperature sensor cannot be used when using the
temperature sensor Autoaccurate sensor.
1.
Unscrew the threaded fi tting on the condenser.
2.
Insert the vapor temperature sensor so in the condenser that the white PTFE side of
the seal (2) points to the glass thread.
3.
Screw the vapor temperature sensor together with the threaded fi tting.
4.
Connect the plug (1) in the socket on the tower to the left of the unit (see chapter 5).
1
3
2
Figure4-30: Temperature sensor Autoaccurate sensor
1
Threaded fi tting cap
3
PTFE side of the sealing tube
2
Plug
Only in conjunction with glassware sets G3 and G6.