Troubleshootimg
45
Error / message
Cause
Remedy
No evacuation
Power switch Rotavac vario
control or Rotavac vario tec is
not activated.
Vacuum valve is defective
➜
Activate the power switch
Fuses defective or blown fuse
➜
Replace fuses (see chapter
“Assembly, Electrical
connection”)
Insufficient vacuum
System leaks
➜
Check seals, tubing and
connections
➜
Check stopcock, grease if
necessary
Vacuum pump is defective
➜
Observe manufacturer‘s
instructions of the vacuum
pump
Unit suddenly
switches off
Timer is programmed
➜
Check and switch off the
timer, if necessary (see
chapter „Operation, Timer
functions”)
Fuses defective or blown fuse
➜
Replace fuses (see chapter
“Assembly, Electrical
connection”)
When pressure is detected,
the Hei-VAP Precision is emer
-
gency stopped.
➜
The Hei-VAP Precision is reset
to the starting condition when
pressing any key.
Check the settings in the
setup menu.
➜
Calibrate the vacuum sensor
(see chapter „Operation”
„More functions”)
Tabelle 7-1: Error codes
Additional conditions
Error / message
Reaction
Auto accurate sensor is
not connected
▪ Change to the SET pressure mode
▪ The current pressure value is accepted as target value,
similar to Hold/Accept in the SET pressure mode
▪ Auto accurate mode cannot be selected
Vapor sensor is
defective
▪ The actual value curve “- - -“ is displayed instead of the
vapor temperature.
The vapor sensor is not
connected
▪ The vapor temperature is not displayed in the actual value
display
Heater can only be
activated at > 0 °C
Table 7-2: Additional conditions