Assembly
65
1
Condenser
3
Tension spring
2
Threaded clamping screw
4
Drive unit
1.
Version glassware sets G1 and G3:
▪
Push the threaded clamping screw (2) over the condenser flange.
▪
Push the spring (3) over the condenser flange.
▪
Attach the condenser (1) to the drive unit (4) with the threaded clamping screw.
2.
Version glassware set G5 (depending on coolant):
▪
Push the threaded clamping screw (2) over the cold trap jacket flange.
▪
Push the spring (3) over the cold trap liner flange.
▪
Screw the cold trap jacket with the threaded clamping screw to the drive unit.
▪
Insert the o-ring in the sealing groove of the cold trap jacket.
▪
Insert the PTFE centering ring in the cold trap jacket and insert the cold trap in-sert
(cold finger).
The flange is suctioned compressed by the vacuum and the system is closed vacuumtight
by the inserted o-ring.
Variante Glassatz G6 (Kühler für Rückflussdestillation):
1
2
3 4 5
6
Illustratiion 4-11: Glassware G6
1
Center piece
4
Gasket G6
2
Threaded connection
5
G6 intermediate part (5)
3
Tension spring
6
G6 threaded coupling
▪
Push the threaded connection (2) over the condenser flange.
▪
Push the spring (3) over the condenser flange.
▪
Screw the G6 (6) gland on the drive head (Figure 4-10 (4).
▪
Insert the G6 intermediate part (5) with seal G6 (4) into the gland G6 (6).
▪
Screw the center piece (1) with the screw (2) on gland G6 (6).
▪
Insert the condenser in the upper cut on the center piece (1).