- 30 -
Solution
:
Please
check if the foot pad adjustment is correct. Under normal circumstances, the front side is
slightly higher than the back side.
Open the door and check if the filter at the bottom of the sterilization chamber is blocked.
None of the above can help reduce the consumption of distilled water, please contact professional
distributor authorized by
HӓTmed.
9.3
The drying effect is not good
Faulty description
:
Drying cannot achieve satisfactory results
Solution
:
Please
check if the foot pad adjustment is correct. Under normal circumstances, the front side is
slightly higher than the back side.
O
pen the door and check if the filter at the bottom of the sterilization chamber is blocked.
Please check if
the load exceeds the maximum loading capacity(especially textiles) and if the load
is placed correctly in the chamber(do not touch the wall of the pressure chamber directly) and if the
standard trays and tray rack are used for loading.
Activate automatic warm‐up.
Open the left side cover and check if the filter is clogged
None of the above can help improve the effect of drying, please contact professional distributor
authorized by
HӓTmed.
9.4
No distilled water
Faulty description
:
No distilled water
Solution
:
Please check if there is water in distilled water tank.
Please check if the distilled water pipe is bent or knotted and inserted into place.
9.5
Unable to open the door
Faulty description
:
The door cannot be opened even if the emergency button is activated.
Solution
:
Please check if the air filter at the rear side of sterilizer is blocked
Take off side cover,
unplug the silicone tube connected to the air filter solenoid valve and discharge
the negative pressure inside the chamber.
When you hear the exhaust sound gradually weaken until it disappears, you can open the door.