- 29 -
Clean the filter at
the bottom of the
sterilization chamber
Sterilize 200 times
Wrench
Unscrew the filter with a wrench and wash it
with clean water.
Clean trays
Sterilize 100 times
Rinse it with clean water, if stained, clean it
with toothpaste
Sterilization chamber
Every week
Wipe the sterilization chamber with a clean
and wet medical cotton cloth
,
then wipe the
chamber with a clean and dry medical cotton
cloth
8.2.2
Regular maintenance
WS310‐2016 and WS506‐2016 require a comprehensive
maintenance of the small steam sterilizer every year to ensure
good equipment working condition and sterilization effect
Maintenance and take care of this device must be performed by a
professional authorized by the manufacturer.
9
Faults and common problem handling
9.1
No display on the screen
Faulty description
:
The display does not light up normally when the sterilizer is turned on.
Solution
:
Check if there is power in the main power outlet
Check if the power plug of this product is plugged in.
The screen still does not appear, please contact professional distributor authorized by
HӓTmed.
9.2
Large consumption of distilled water
Faulty description
:
Distilled water consumption is much larger than the specified amount