- 21 -
5.3
Helix Test
The Helix test is used to verify the steam penetration and saturation for A type hollow instruments, such as
dental handpieces. Select the Helix test in the test program interface (Figure 5.1).
Test methods and test results are judged in accordance with the
information and requirements provided by the test package
manufacturer.
5.4
Drainage program
This product allows users to perform drainage procedures when needed. This procedure is used to drain water
remaining in a small steam sterilizer.
When the water in the chamber is obvious after opening the door,
it is recommended to use the drainage program before loading the
item for sterilization. Especially when the program is abnormally
stopped, this situation is easy to occur. Please open the door for
inspection.
Please carry out the drainage program before handling this
product.
6
System setting
Enter the system settings interface (Figure 6.1) for time, date, language, warm‐up, automatic printing,
additional drying settings, and system reset.