4.5
Manual stop
A running program can be manually stopped at any stage. If the sterilization process is stopped before
the start of drying, the loaded items are not sterilized. If the sterilization procedure is manually stopped after the
start of drying, the loaded items have been effectively sterilized, but not fully dried. In the case of fully loaded
packages, the possibility of wet bags is very high and these items cannot be stored because Because aseptic storage
places high demands on the dryness of the packaged items.
After the program is manually stopped, there may be high temperature
steam coming out and causing burns when the door is opened
!
Do
not touch the product, loaded items, chamber or door cover
with bare hands.
Be sure to use the tray handle to take the loaded items
.
Manually stopping the sterilization process at an earlier stage may cause an
infection
!
If the sterilization process is stopped before the start of drying, the
loaded items are not sterilized
,
The use of these items may
endanger the health of the operator and the patient.
If necessary, the loaded items should be repackaged and sterilized
.
If you need to
manually stop the current program, please press
(Fig 4.3). After manual stop, the screen
displays the program end interface and manually stops the fault information. The intelligent controller is processed
according to the current situation to achieve a safe pressure inside the chamber and other parts of the device also
ensure a safe state. After the device is in a safe state, the controller is allowed to operate the door open button,
the required time and the program running state when manually stopped, usually takes several minutes.
134 program
P: 0 Bar T: 25.0
℃
Door status
00:00:00
Press
to stop
the program
Program stop
The program stops in
progress
…
P: 0.85Bar