- 5 -
Working environment
pressure
70 Kpa ~ 106 Kpa
Working temperature
5
o
C ~ 40
o
C
Storage and transportation
environment temperature
‐20
o
C ~ 55
o
C
environment humidity
≤ 85%
Maximum wastewater
temperature
100
o
C
Maximum noise
70 dB(A)
Electromagnetic
compatibility
Compliance with GB/T18268.1‐2010;
Works properly in its electromagnetic
environment, but does not constitute an unacceptable electromagnetic
harassment of anything in the environment
。
At the same time, it has the
ability to resist electromagnetic interference.
According to GB 4824‐2013
,
Electromagnetic radiation meets Group 1 Class
B requirements
.
Standards compliant
YY/T 0646‐2015; GB4793.1‐2007; GB4793.4‐2001;
Note
1
:
The maximum operating voltage range
:
198V ~ 242V
Note
2:
Anti‐electromagnetic interference is the ability of the sterilizer to
work properly in the presence of electromagnetic interference
1.4
Loading method
The tray rack is included in the delivery of sterilizer. The space inside the tray rack is a space that can be
effectively sterilized, always use the tray rack when loading items to be sterilized. The tray rack can hold 5 trays or
adapt to thickness of different loads or containers after rotating 90
o
C.
Fig 1.6 Loading method
1.5
Intelligent controller and control panel
Intelligent controller is the core of small steam sterilizer, In addition to completing operations such as rapid
acquisition, processing, fault detection, and control of the system's real‐time parameters, the controller also
intuitively plots the temperature and pressure versus time curve in the chamber.
The intelligent controller is easy and efficient to operate with LCD screen with 4 buttons.