- 9 -
2.4
Connection requirements
2.4.1
Electrical connections
220V/50Hz power supply. The maximum power supply
fluctuation range is 198V ~ 242V.
Use a separate 16A power circuit. Provides 30mA leakage
protection.
Make sure the power cable is not knotted or pressed
.
Do not use force to pull the power cable,
Do not place the power cable near a heat source.
Do not touch the power cable with a sharp object.
Do not replace the power cable or power plug. Do not use a
damaged power cable or power plug
Do not use this device if the power cable or power plug is
damaged. The power cable or power plug must be replaced by a
professional authorized by the manufacturer of this device.
2.4.2
Distilled water injection
Open up cover of the internal water tank and Inject the distilled water into the internal water tank through the
water tank deflector, please be noted that the distilled water added should not exceed the height of the water
deflector, generally stored in a standard 12L distilled water tank.Full tank of distilled water can complete up to 20
sterilizations. Please pay attention to the amount of distilled water remaining in the tank to avoid affecting the
normal operation of this device.
Compliance with YY/T 0646‐2015
high quality distilled water as
specified in appendix B
or
demineralized water
because the degree
of formation of rust stains depends on whether the water quality of
the distilled water is up to standard.
Conductivity ≤20μs/cm
;
If the inlet pressure is
>30 Kpa m
ay cause excess distilled water to
enter the chamber and affect the operation of the device.
In order to prevent the distilled water from being contaminated, the
lid of the distilled water tank must be tightened or tight at all times.
For small steam sterilizer with internal water tank, please drain the
waste water tank before injecting distilled water
.
Regularly clean the distilled water tank at least once a month
.