- 7 -
3.Holding
After the temperature and pressure reached the set value, holding phase
begins and maintains stable temperature and pressure for a period of time
to ensure that the sterilization effect meets the requirements of the
specification
4.Depressurizatio
n
The pressure in the chamber is released after the holding time is over. At the
same time, the steam in the steam generator is discharged.
5.Drying
After sterilization, the sterilized items are dried. Drying begins after pressure
is released. Under the effect of a high efficiency air filter, the air in the
chamber is always sterile during the drying process
6.Intake
The drying process is also the process of intake, make sure that the pressure
in the chamber returns to environmental pressure at the end of the
program.
1.7.2
Vacuum test
The purpose of the vacuum test is to test the leak rate of the sterilizer. No sterilization of the items in the
chamber during the process. Please run the vacuum test in a cold and dry chamber without loads.
Process name
Description
1.Vacuum
The air in the chamber is evacuated till the pressure
in the chamber reaches
the set value of the vacuum test
2.Balancing time
After waiting for 5 minutes, the vacuum everywhere in the chamber tends to
be uniform.
3. Test holding
time
Test holding time is 10 minutes,
The device continuously monitors the
pressure in the chamber during the period
4.End of test
After the test is over, the screen displays the test results: vacuum and leak
rate
.
2
Installation and handling instructions
2.1
Unpacking precautions
Installation, setup, and commissioning should be performed by a
professional authorized by the manufacturer of this product.
This product is taken out of the box and transported must be
operated by 2 people.
Total weight
:
Q50B 59.6 kg; Q51B 61.6 kg;
Q70B 64.8 kg; Q71B 66.8 kg