- 17 -
Fig 4.3 Manual stop
4.6
Sterilization procedure completed
When the sterilization process ends normally, the program end interface and program running success
information are displayed on the screen. If the automatic printing function is set and the printer is connected, the
product will automatically print the sterilization record. Please press
to open the door.
Careful burns
!
Do not
force the door to open. Failure to do so may result in
damage to the product and/or hot steam coming out of the
chamber.
After the sterilization is finished, wait for a while until the sterilizer
is fully cooled before opening the door.
Do not touch any metal surface
.
Do not touch the sterilized loaded items, chamber and door covers
with bare hands to avoid being burned.
Be sure to use the tray handle to remove the loaded items.
Balance
when removing the tray to prevent the contents of the
tray from slipping
.
Damaged packaging devices or items
may not be effectively sterilized,
and exposure to these items may endanger the health of the operator or
patient.
If damaged packaging if found, please repackage the item.
Waiting for 5 minutes after sterilization before removing the
loaded items, which is good for drying the sterile items.
The
11th edition (2017) of the correct handling method for
medical device cleaning and disinfection issued by the
international authority AKI, Germany. It is recommended that the
wet bag be judged as having water drops remaining on the
package or in the package after 15 minutes of sterilization.
Physical parameters should be monitored for each sterilization
cycle, record and check process variables. When the sterilization
temperature set in this cycle is 121 ° C, the holding time should be
no less than 15 minutes. When the sterilization temperature set in