- 19 -
5
Functional testing and other procedures
5.1
Vacuum test
When first installed, unused for a long time, moved, or periodically tested,vacuum test is recommended to
detect leakage in the system of this sterilizer:
Fig 5.1 Vacuum Test
Vacuum test should be carried out when the system is cooled and
the chamber is dry.
In order to successfully complete the vacuum test,please check
that the door of sterilizer is closed and the cooling water valve is
open before starting the program.
1 –Sterilization program
2 – Test program
3 – System setting
4 – File record
Vacuum test program
B&D test program
HELIX test program
press
to select
07/25/2017 09:23:58
Vacuum test program
15Min
T: 25.0
℃
P: 0.00 Bar
Door status
press
to enter
press
to start up
Vacuum test program
P: 0.00 Bar T: 25.0
℃
Door status
00:00:00